Calibration procedure
To carry on the recalibration the following instruments are needed:
- Known concentration gas cylinder suitable for the sensor, complete with a pressure regulator
- Flow meter
- Hose with Tee fitting to connect the cylinder to the flow meter and to the instrument
In the following is described a recalibration example for the CO sensor.
→
1. Start the instrument
ATTENTION
•
Make sure autozero is performed in clean air and completes properly.
•
Do not connect the gas probe to the instrument.
•
Check the battery charge level or connect the power adapter to avoid data loss
during recalibration.
07/08/14
10:00
Eff. tot
%
91.4
Loss tot
%
8.6
∆T
°C
74.7
T air
°C
15.4
T flue
°C
190.1
Ex. Air
1.25
CO
2
%
9.3
O
2
%
4.2
Combustion analysis
07/08/14
10:00
Eff. tot
%
91.4
Loss tot
%
8.6
∆T
°C
74.7
T air
°C
15.4
T flue
°C
190.1
Ex. Air
1.25
CO
2
%
9.3
O
2
%
4.2
Combustion analysis
Keep
pressed for
a few
seconds
2. Once autozero is completed press the key and select the diagnostic icon.
07/08/14
10:00
Sensors
Hardware
Gas probe
Pump
Diagnostic
On site cal.
Password
On site calibration
07/08/14
10:00
0 0 0 0
OK
Password
On site calibration
07/08/14
10:00
1 1 1 1
OK
OK
OK
Enter the
recalibration menu
password
1111.
Serial number: 1000
Firmware version: 1.00
E4500-N
Содержание E4500
Страница 2: ......
Страница 20: ...20 K000000000EJ 023826A0 100714 7 4 Connection diagram AAC SO01 AA SF A AAC TA03A AAC KP01 AA AL05 AA SA08...
Страница 109: ...109 K000000000EJ 023826A0 100714 12 11 Measurements External probe Not available...
Страница 138: ......
Страница 140: ......
Страница 141: ...NOTE...
Страница 142: ...NOTE...
Страница 143: ......