User Manual EE355 Dew Point Sensor

6.2  Modbus Register Map

The measured data is saved as 32 bit floating point values (data type FLOAT) and as 16 bit signed integer 

values (data type INTEGER). Depending on the measurement unit selected, the measurements are 

saved in SI or US/GB units. Measurement unit selection is made by means of the ordering code or with 

the help of the EE-PCS Product Configuration Software.

FLOAT 32 bit:

Parameter

SI Unit US/GB Unit Register number

1)

 [Dec] Register address

2) 

[HEX]

Read register: function code 0x03 / 0x04

Temperature T 

°C

°F

26

0x19

Relative humidity RH, Uw 

% RH

%

28

0x1B

Water vapout partial pressure e’

mbar

psi

30

0x1D

Dew point temperature Td 

°C

°F

32

0x1F

Frost point temperature Tf 

°C

°F

42

0x29

Volume concentration Wv 

ppm

ppm

60

0x3B

INTEGER 16 bit:

Parameter

SI Unit US/GB Unit Register number

1)

 [Dec] Register address

2) 

[HEX]

Read register: function code 0x03 / 0x04

State of auto-calibration

0 ... normal operation

1 ... auto-calibration active

501

0x1F4

Write register: function code 0x06

Modbus address (Slave ID)

1

0x00

RS485 Data Transmission

2

0x01

1) Register number starts from 1

2) Register address starts from 0

7  Maintenance

7.1  Filter Replacement

A contaminated filter cap should replaced by a new one. The order number for new filters is HA010103.

When replacing the filter, please note the following:

 

ƒ

Unscrew the filter cap very carefully to avoid damaging the sensing element.

 

ƒ

Handling the filter might clog the pores. Use gloves to screw on the new filter.

Filter cap

Sensing element

Fig. 7 

EE355 Filter and Sensor

7.2  Cleaning the Sensing Element

Caution:

 

ƒ

Never touch the sensing elements.

 

ƒ

Any attempt to clean the sensing elements mechanically such as rubbing or brushing leads certainly 

to their irreversible damage.

Please observe the E+E “Cleaning Instructions”.

Содержание EE355

Страница 1: ...BA_EE355 v1 4 Modification rights reserved User Manual EE355 Dew Point Sensor down to 60 C Td...

Страница 2: ...equipment has been tested and found to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmf...

Страница 3: ...3 4 Auto calibration 7 3 5 Measuring Range and Accuracy 7 4 Installation 8 4 1 Installation Location 8 4 2 Installing directly into the Process 8 4 3 Installing the Sensor in a Sampling System 8 5 Set...

Страница 4: ...ioning shall be performed by qualified personnel only Use the EE355 only as intended and observe all technical specifications Do not use EE355 in explosive atmosphere or for measurement of aggressive...

Страница 5: ...protect them from accidental commissioning Mark the damaged product as defective A faulty device may only be investigated and possibly repaired by qualified trained and authorized staff If the fault...

Страница 6: ...3 74 2 9 30 1 2 12 0 5 G1 2 1 2 NPT 92 3 6 34 5 1 36 89 3 5 AD27 AD22 Fig 1 EE355 Dimensions G ISO Thread in mm inch 140 5 5 77 3 74 2 9 30 1 2 12 0 5 G1 2 1 2 NPT 92 3 6 34 5 1 36 89 3 5 AD27 AD22 F...

Страница 7: ...ed up to 5 times during the first hour of operation During auto calibration the analogue output can switch to one of two states Frozen output signal keeping the last measured value default behaviour o...

Страница 8: ...ensor is protected by a cap that keeps the dew point sensor dry The cap should only be removed right before installation into the application Fig 4 Remove protection cap 4 2 Installing directly into t...

Страница 9: ...ials give off moisture over time when flushed with very dry gas after being at ambient humidity 1 2 4 3 5 t h 70 60 50 40 30 20 Td C Fig 6 Moisture given off by different tubing materials National Phy...

Страница 10: ...ction the power supply must be strong enough to ensure supply voltage within the specified range see technical data at any time and at all devices in the bus This is particularly relevant when using l...

Страница 11: ...Unit US GB Unit Register number1 Dec Register address2 HEX Read register function code 0x03 0x04 State of auto calibration 0 normal operation 1 auto calibration active 501 0x1F4 Write register functio...

Страница 12: ...ownload www epluse com configurator EE PCS Connection cable 5 pole M12x1 socket free ends PUR insulation 1 5 m 4 9 ft 5 m 16 4 ft 10 m 32 8 ft HA010819 HA010820 HA010821 Sampling cell G 1 2 with quick...

Страница 13: ...inless steel 1 4404 AISI 316L IP65 NEMA 4 x Electrical connection M12x1 5 poles stainless steel 1 4404 Filter Stainless steel sintered Operating conditions 40 70 C 40 158 F 0 100 RH Storage conditions...

Страница 14: ...lin 69130 Ecully Tel 33 4 74 72 35 82 E mail info epluse fr E E Elektronik Germany Sch ne Aussicht 8 C 61348 Bad Homburg Tel 49 6172 13881 0 E mail info epluse de E E Elektronik Italy Via Alghero 17 1...

Отзывы: