User Manual EE355 Dew Point Sensor

1  General

This user manual serves for ensuring proper handling and optimal functioning of the device. The user 

manual shall be read before commissioning the equipment and it shall be provided to all staff involved 

in transport, installation, operation, maintenance and repair. The user manual may not be used for the 

purposes of competition without the written consent of E+E Elektronik® and may not be forwarded to 

third parties. Copies may be made for internal purposes. All information, technical data and diagrams 

included in these instructions are based on the information available at the time of writing.

1.1  Explanation of Symbols

This symbol indicates safety information.

It is essential that all safety information is strictly observed. Failure to comply with this information can 

lead to personal injuries or damage to property. E+E Elektronik® assumes no liability if this happens.

This symbol indicates instructions.

The instructions shall be observed in order to achieve optimal performance of the device.

1.2  Safety Instructions

1.2.1  General Safety Instructions

 

ƒ

Avoid any unnecessary mechanical stress and inappropriate use.

 

ƒ

When replacing the filter cap make sure not to touch the sensing elements.

 

ƒ

The device must be operated with the filter cap on at all times.

 

ƒ

For sensor cleaning and filter cap replacement please see the “Cleaning Instructions” at  

www.epluse.com.

 

ƒ

Installation, electrical connection, maintenance and commissioning shall be performed by qualified 

personnel only.

 

ƒ

Use the EE355 only as intended and observe all technical specifications.

 

ƒ

Do not use EE355 in explosive atmosphere or for measurement of aggressive gases.

 

ƒ

Do not apply the nominal voltage to the data lines.

1.2.2  Intended Use

The EE355 is intended for the dew point temperature (Td) measurement of compressed air and other 

non-corrosive and non-flammable gases at pressures. The sensor can be installed in a pressurized 

system up to 80 bar (1 160 psi). The probe is pressure-tight and has leak rate B according to EN 12266-1.

The use of the EE355 in any other way than described in this manual bears a safety risk for people and 

the entire measurement installation and is therefore not allowed. 

The manufacturer cannot be hold responsible for damages as a result of incorrect handling, installation, 

and maintenance of the equipment.

In order to avoid damage to the instrument or health hazards, the measuring equipment must never be 

manipulated with tools that are not specifically described in this manual.

The sensor may only be utilized in accordance with the conditions defined in the technical data. 

Otherwise, inaccuracies of the measurement will occur and equipment failures cannot be ruled out.

The steps recommended by the manufacturer for installation, inspections and maintenance work must 

be observed and carried out for the safety of the user and for the functionality of the equipment.

Unauthorized product modification lead to loss of all warranty claims. This may be accomplished only 

with an explicit permission of E+E Elektronik

®

!

The device is constructed for the operation of separated extra-low voltage (SELV).

Содержание EE355

Страница 1: ...BA_EE355 v1 4 Modification rights reserved User Manual EE355 Dew Point Sensor down to 60 C Td...

Страница 2: ...equipment has been tested and found to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmf...

Страница 3: ...3 4 Auto calibration 7 3 5 Measuring Range and Accuracy 7 4 Installation 8 4 1 Installation Location 8 4 2 Installing directly into the Process 8 4 3 Installing the Sensor in a Sampling System 8 5 Set...

Страница 4: ...ioning shall be performed by qualified personnel only Use the EE355 only as intended and observe all technical specifications Do not use EE355 in explosive atmosphere or for measurement of aggressive...

Страница 5: ...protect them from accidental commissioning Mark the damaged product as defective A faulty device may only be investigated and possibly repaired by qualified trained and authorized staff If the fault...

Страница 6: ...3 74 2 9 30 1 2 12 0 5 G1 2 1 2 NPT 92 3 6 34 5 1 36 89 3 5 AD27 AD22 Fig 1 EE355 Dimensions G ISO Thread in mm inch 140 5 5 77 3 74 2 9 30 1 2 12 0 5 G1 2 1 2 NPT 92 3 6 34 5 1 36 89 3 5 AD27 AD22 F...

Страница 7: ...ed up to 5 times during the first hour of operation During auto calibration the analogue output can switch to one of two states Frozen output signal keeping the last measured value default behaviour o...

Страница 8: ...ensor is protected by a cap that keeps the dew point sensor dry The cap should only be removed right before installation into the application Fig 4 Remove protection cap 4 2 Installing directly into t...

Страница 9: ...ials give off moisture over time when flushed with very dry gas after being at ambient humidity 1 2 4 3 5 t h 70 60 50 40 30 20 Td C Fig 6 Moisture given off by different tubing materials National Phy...

Страница 10: ...ction the power supply must be strong enough to ensure supply voltage within the specified range see technical data at any time and at all devices in the bus This is particularly relevant when using l...

Страница 11: ...Unit US GB Unit Register number1 Dec Register address2 HEX Read register function code 0x03 0x04 State of auto calibration 0 normal operation 1 auto calibration active 501 0x1F4 Write register functio...

Страница 12: ...ownload www epluse com configurator EE PCS Connection cable 5 pole M12x1 socket free ends PUR insulation 1 5 m 4 9 ft 5 m 16 4 ft 10 m 32 8 ft HA010819 HA010820 HA010821 Sampling cell G 1 2 with quick...

Страница 13: ...inless steel 1 4404 AISI 316L IP65 NEMA 4 x Electrical connection M12x1 5 poles stainless steel 1 4404 Filter Stainless steel sintered Operating conditions 40 70 C 40 158 F 0 100 RH Storage conditions...

Страница 14: ...lin 69130 Ecully Tel 33 4 74 72 35 82 E mail info epluse fr E E Elektronik Germany Sch ne Aussicht 8 C 61348 Bad Homburg Tel 49 6172 13881 0 E mail info epluse de E E Elektronik Italy Via Alghero 17 1...

Отзывы: