Operating instructions for EE310 Humidity / Temperature Transmitter
6.4 RH / T adjustment and calibration
Humidity calibration and adjustment
Depending on the application and the requirements of certain industries, there might arise the
need for periodical humidity calibration (comparison with a reference) or adjustment (bringing the
device in line with a reference).
Calibration and adjustment at E+E Elektronik
Calibration and/or adjustment can be performed in the E+E Elektronik calibration laboratory. For
information on the E+E capabilities in ISO or accredited calibration please see
www.eplusecal.com
.
Calibration and adjustment by the user
Depending on the level of accuracy required, the humidity reference can be:
•
Humidity Calibrator (e.g. Humor 20), please see
www.epluse.com/humor20
.
•
Handheld device (e.g. Omniport30), please see
www.epluse.com/omniport3
0.
•
Humidity calibration Kit (e.g. E+E Humidity Standards), please see
www.epluse.com/EE310
.
Perform 1 or 2 point adjustment via EE-PCS Product Configuration Software or via display
(details see below).
Menu
Customer
adjustment
Offset
Reference value
Current
measured value
Calibration status
i
Reset to factory calibration
2 point adjustment
Reference
value low
Reference value
high
Rel. humidity
Temperature
value
value
Reference value
Current measured
value
Fig. 27
Adjustment menu
measurement value
Measured value
Displayed value
reference value
Offset adjustment point
FACT
ORY
/ CURRENT
ADJUSTMENT
NEW
ADJUSTMENT
Fig. 28
Offset adjustment