User Manual EE1900 Humidity Module for Climate Chambers

4.3  Error Indication on the Analogue Output (NAMUR)

The EE1900 features an error indication on the analogue output according to NAMUR  

recommendations. A device failure is indicated by 11 V in case of voltage output and by 21 mA 

in case of current output. 

Important: 

Upon delivery the NAMUR error indication is disabled. It can be enabled with the free EE-PCS Product 

Configuration Software.

5  Automatic Sensor ReCovery (ARC) 

Certain chemical pollution may lead to drift of the sensing element. Using the sensor recovery function, 

the contamination is removed by a controlled heating of the sensor. After the heating process, the sensor 

quickly returns to normal measuring conditions (refer to Fig. 5).

Start:

The ARC can be started either manually by pressing the “Sensor Recovery” button (No. 6 - Fig. 3) for 

min. 3 seconds or automatically by applying a voltage of 5 - 30 V DC to the potential free input (No. 3 

“Heat IN“ - Fig. 3) for at least 3 seconds. 

Abortion:

To cancel the ARC process, the same procedure is valid as to starting it: pressing the “Sensor 

Recovery” button (No. 6 - Fig. 3) or a voltage of 5 - 30 V DC at the “Heat IN“ interface for min. 3 

seconds. 

Duration:

A sensor recovery cycle takes around 23 minutes, of which 20 minutes heat-up time, followed by 3 

minutes cooling time. The cooling time is necessary also in case the ARC process is cancelled.

Current consumption in ARC mode

Supply

Consumption

15 V DC

~120 mA (max.)

24 V DC

~80 mA (max.)

35 V DC

~55 mA (max.)

17 V AC 

~210 mA

rms 

(max.)

24 V AC

~160 mA

rms  

(max.)

29 V AC

~140 mA

rms  

(max.)

+
GND

Power supply

5 - 30 V DC

+
GND

GND

GND

Pulldown: 

100 k

Power supply 

5 - 30 V DC

in

out

ARCStatus

V

Galvanically separated input/output

Fig. 4 

ARC trigger and status connection

Important:

ARC can only be started if RH < 85%.

Содержание EE1900

Страница 1: ...BA_EE1900 v1 1 Modification rights reserved User Manual EE1900 HUMIDITY MODULE FOR CLIMATE CHAMBERS...

Страница 2: ...equipment has been tested and found to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmf...

Страница 3: ...al 5 2 1 1 Sensing Probes 6 2 1 2 Filter Caps 6 2 2 Dimensions 6 2 3 Installation 6 2 3 1 Electronics Board 6 2 3 2 Sensing Probe 6 3 Electrical Connection 7 4 Setup and Configuration 8 4 1 Selection...

Страница 4: ...commissioning shall be performed by qualified personnel only 1 2 2 Mounting Start up and Operation The device has been produced under state of the art manufacturing conditions has been thoroughly test...

Страница 5: ...E E Elektronik are developed and manufactured observing of all relevant requirements with respect to environment protection Please observe local regulations for the device disposal For disposal the i...

Страница 6: ...47 Probe length see ordering guide Cable length see ordering guide Filter length see table below 3 5 0 14 3 5 0 14 Fig 1 Dimensions of EE1900 2 3 Installation ESD handling regulations must be strictly...

Страница 7: ...tor If the probe is installed into a partition wall the probe s backend shall be isolated thermally 3 Electrical Connection Supply Fig 2 Power supply 15 35 V DC 17 29 V AC Connection supply and output...

Страница 8: ...RH and error indication according NAMUR see chapter 4 3 is disabled After switching the output between voltage and current with the slide switch the scaling of the output can be set using the E E Pro...

Страница 9: ...manually by pressing the Sensor Recovery button No 6 Fig 3 for min 3 seconds or automatically by applying a voltage of 5 30 V DC to the potential free input No 3 Heat IN Fig 3 for at least 3 seconds A...

Страница 10: ...overy The ARC mode is indicated by the closed contact at the optocoupler output No 3 Status Fig 3 and by the red LED flashing very quickly The optocoupler output thus indicates wether the data on the...

Страница 11: ...guration PC HA011017 EE1900 The adjustment shall be performed for the RH output against an appropriate RH reference See Calibration Kit User Guide at www epluse com ee1900 7 Self Diagnosis and Error M...

Страница 12: ...RH Dew point Td Working range 20 80 C 4 176 F Accuracy 2 C 3 6 F for Tambient Td 20 C 36 F General Response time RH t10 90 at 20 C 68 F typ 15 s with stainless steel grid filter2 Output signal 0 1 5...

Страница 13: ...lin 69130 Ecully Tel 33 4 74 72 35 82 E mail info epluse fr E E Elektronik Germany Sch ne Aussicht 8 C 61348 Bad Homburg Tel 49 6172 13881 0 E mail info epluse de E E Elektronik Italy Via Alghero 17 1...

Отзывы: