User Manual EE074 Temperature Probe with Modbus RTU
6.3 Configuration Interface
The EE074 is ready to use and does not require any configuration by the user. The user can change
the factory setup by using the Modbus configuration adapter (order code HA011018) and the EE-PCS
Product Configuration Software.
One can change the digital communication settings and perform a measurement adjustment in the form
of an offset or as a 2-point adjustment.
EE-PCS PRODUCT CONFIGURATION SOFTWARE
1. Download the EE-PCS Product Configuration Software from
install it on the PC.
2. Connect the E+E device to the PC using the appropriate configuration cable.
3. Start the EE-PCS software.
4. Follow the instructions on the EE-PCS opening page for scanning the ports and identifying the
connected device.
5. Click on the desired setup or adjustment mode from the main EE-PCS menu on the left and follow
the online instructions of the EE-PCS.
PC
HA011018
6.4 T Adjustment and Calibration
6.4.1 Temperature Adjustment and Calibration
Definitions:
Adjustment: the specimen is brought in line with the reference.
Calibration: the specimen is compared with a reference and its deviation from the reference is docu-
mented.
Depending on the application and the requirements of certain industries, there might arise the need for
periodical temperature calibration or adjustment
6.4.2 Calibration and Adjustment at E+E Elektronik
Calibration and/or adjustment can be performed in the E+E Elektronik calibration laboratory. For infor-
mation on the E+E capabilities in ISO or accredited calibration please see
.
6.4.3 Calibration and Adjustment by the User
Depending on the level of accuracy required, the temperature reference can be:
• Liquid bath calibrator
• Dry block calibrator
• Climate camber