E+E Elektronik EE074 Скачать руководство пользователя страница 8

8

User Manual EE074 Temperature Probe with Modbus RTU

5.3  Modbus RTU Example

Example of MODBUS RTU command for reading the temperature (FLOAT value) T = 26,953624 °C 

from the register 0x3EA

Device EE074; slave ID 233 [0xE9]

Decoding of floating point values:

Floating point values are stored according IEEE754 standard. The byte pairs 1, 2 and 3, 4 are inverted 

as follows:

Example:

Please note:

 For obtaining the correct value, both registers have to be read within the same reading 

cycle. The measured value can change between two Modbus requests, therefore, exponent and  

mantissa may get inconsistent.

For more details, please see the Modbus Application Note available at 

https://www.epluse.com/EE074

 

and chapter 6.3 of the Modbus reference document 

http://www.modbus.org/docs/Modbus_Applicati-

on_Protocol_V1_1b3.pdf

.

6  Maintenance and Service

EE074 does not require any special maintenance, nevertheless for high accurate measurements 

especially over wide T ranges it is recommended to calibrate the probe every 12 months. If needed, the 

enclosure may be cleaned and the device may be re-adjusted as described below.

6.1  Cleaning

Use a damp soft cloth to remove deposits of dust or dirt from the exterior of the probe. Do not use any 

solvents or abrasive cleaning agents.

6.2  Repairs

Repairs may be carried out by the manufacturer only. The attempt of unauthorized repair excludes any 

warranty claims.

Modbus ID 

address

Function 

code

Starting 

address Hi

Starting 

address Lo

No. of 

register Hi

No. of 

register Lo

CRC

Request [Hex]:

E9

03

03

EA

00

02

F2

93

Modbus ID 

address

Function 

code

Byte count Register 1 

value Hi

Register 1 

value Lo

Register 2 

value Hi

Register 2 

value Lo

CRC

Response [Hex]:

E9

03

04

A1

06

41

D7

21

CE

MMMMMMMM

MMMMMMMM

SEEEEEEE

EMMMMMMM

Byte 3

Byte 4

Byte 1

Byte 2

Response [Hex]

Value in decimal

 Byte 1 (Register 2 - Hi)

Byte 2 (Register 2 – Lo)

Byte 3 (Register 1 - Hi)

Byte 4 (Register 1 - Lo)

41

D7

A1

 06

26.953624

Содержание EE074

Страница 1: ...BA_EE074_e V1 0 technical data are subject to change User Manual EE074 Temperature Probe with Modbus RTU...

Страница 2: ...equipment has been tested and found to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmf...

Страница 3: ...5 4 2 Duct Mount 6 4 3 Immersion Well Mount 6 4 4 Recommendations for Accurate Temperature Measurement 6 5 Modbus Map 7 5 1 Modbus Setup 7 5 2 Modbus Register Map 7 5 3 Modbus RTU Example 8 6 Maintena...

Страница 4: ...facturing conditions has been thoroughly tested and has left the factory fulfilling all safety criteria The manufac turer has taken all precautions to ensure safe operation of the device The user must...

Страница 5: ...in mm inch 82 3 23 L 12 0 47 6 0 24 L in mm L in Inch 70 2 81 155 6 16 305 12 07 3 2 Connectivity 4 Installation 4 1 Wall and Ceiling Mount Best measurement performance is achieved when the entire pr...

Страница 6: ...4 Recommendations for Accurate Temperature Measurement Air Temperature Measurement In air best accuracy is achieved with an air flow of min 0 2 m s surrounding the sample In case the EE074 probe is no...

Страница 7: ...dbus protocol settings see Application Note Modbus AN0103 available on www epluse com EE074 INFO read register Parameter Register number1 DEC Protocol address2 HEX Read only register function code 0x0...

Страница 8: ...any special maintenance nevertheless for high accurate measurements especially over wide T ranges it is recommended to calibrate the probe every 12 months If needed the enclosure may be cleaned and t...

Страница 9: ...ed setup or adjustment mode from the main EE PCS menu on the left and follow the online instructions of the EE PCS PC HA011018 6 4 T Adjustment and Calibration 6 4 1 Temperature Adjustment and Calibra...

Страница 10: ...ectronics 40 80 C 40 176 F 0 100 RH probe 70 155 mm 40 80 C 40 176 F 0 100 RH probe 305 mm 70 105 C 94 221 F 0 100 RH Storage conditions 40 80 C 40 176 F 0 90 RH Immersion well Material brass nickel p...

Страница 11: ...0 6460 Office Stuttgart Tel 49 151 538 37 500 E E Elektronik Italy info epluse it Tel 39 02 2707 86 36 E E Elektronik France info epluse fr Tel 33 4 74 72 35 82 E E Elektronik USA office epluse com Of...

Отзывы: