DYNISCO UPR800 Series Instruction Manual
Page | 42
To change or view the Main Analog Output Filter, press the
FUNC
key until
nonE
and
GROUP
show on the display. Press the
t
key until
2
shows in the upper display. Press
the
FUNC
key until the lower display changes to
MO.FL
. Using the
u
or
t
keys, select
the amount of filtering desired, from none (
OFF
) to five seconds. When finished, press
the
FUNC
key to lock in the value and advance to the next parameter. Similarly the
Secondary Analog Output Filter (
SO.FL
) can be set in the same manner.
6.5 Setting the Security Codes:
The security code setting is configured by software (front key board or remotely). There
are three Security levels. When each level has been assigned an access code, the
parameters will be available as follows:
•
Level A: Allows access to parameters in Group 1 Only
•
Level B: Allows access to parameters in Groups 1 and 2 Only
•
Level C: Allows access to parameters to all Groups 1 - 5
Before the unit security levels can be set, the user must first place the instrument in
security setting mode by doing the following: With the unit powered on, press and hold
the
FUNC
key until
GOTO
appears on the lower screen. Press the
t
key until
CODE
displays on the upper screen. Press the
FUNC
key and the unit will reboot into
CODE
mode. Press the
t
key to scroll to the desired security code level to change the value as
needed (see Setting the Security Code below).
Press
FUNC
key to store the change.
To exit the security setting mode, return to the
CODE
by pressing the
FUNC
key. Next
press and hold the
FUNC
key until
GOTO
appears on the lower screen. You will see
CODE.A
on the lower screen, but continue to press the
FUNC
key. When CODE is
displayed press the
u
or
t
keys until
OPr
is displayed. Press the
FUNC
key to return to
Operate mode. The instrument will reboot and return to normal display.
8.7.1 Setting the Security Codes for Level A
To view or change the security code, press and hold the
FUNC
key for approximately 4
seconds. The Lower display changes to GOTO and the upper display reads OPr. If no
buttons are pressed, the indicator reverts back to reading pressure.Press the
t
key until
the upper display reads “CodE”. Press the
FUNC
key and the lower display changes to
CODE.A
. The upper display shows
0,
which indicates no security, and 1 means all
parameters related to levels A, B, and C are always locked). Press the
u
or
t
keys until
the desired security code number (from 2 to 250) appears in the upper display. Press the
FUNC
key to lock in the value. The upper display changes to
1
, and the lower display
changes to
CODE.B
. This means that ONLY Levels B and C are locked, NOT Level A.
8.7.2 Setting the Security Codes for Level B
If you first set
CODE.A
, the lower display will read
CODE.B
; if not, press the
FUNC
key to
move to
CODE.B
. The upper display shows
0,
which indicates no security; or it may
show
1
, which means all parameters related to levels A, B, and C are always locked).
Press the
u
or
t
keys until the desired security code number (from 251 to 500) appears
in the upper display. Press the
FUNC
key to lock in the value. The lower display