DYNISCO SC170 Manual
7
77
7
7
7
77
7
7
2.3.1 Output Terminals
The output terminals 9 and 10 are connected to a power supply
having a nominal 24 Volt DC voltage and capable of supplying 100mA
for the SC170. In the analog mode, these signal conditioners require a
maximum of 23mA. The + OUT and - OUT terminals of the power
supply. The maximum series resistance in the circuit (including wiring
lead resistance) can be calculated using the formula:
Vs - 12
Rs = ————————
0.023
The following chart gives maximum series resistance:
2.3.2 Case Ground
Terminal 8 provides a connection to the metal enclosure of the signal
conditioner. For safety, optimum performance and EMI immunity the
case of the instrument should be connected to a good local earth
ground. When using grounded sensors that are connected to the
local electrical ground, then the signal conditioner case should be
connected to that same ground point.
Max. Series Resistance Rs
Supply Voltage Vs
1300 ohms
42.0 Volts
520 ohms
24.0 Volts
417 ohms
21.6 Volts
250 ohms
18.0 Volts
0 ohms
12.0 Volts