6.0 SPECIFICATIONS
INPUT TYPE & RANGE:
DC Milliamp
-5.0 to +60.0 mA
DC Millivolt
-15.0 to +160 mV
DC Volt
-0.5 to +10.0 V
Potentiometer, 3 wire
0.0 to 10.0 Kohm
Resistance, 2or 3 wire
0.0 to 10.0 Kohm
INPUT RESISTANCE:
DC Milliamp
2.75 Ohms
DC Millivolt
>10M Ohms
DC Volt
1.0M Ohms
ISOLATION: 850 VDC or peak AC
LINEARIZATION:
Square, square root, log, and in to +/- 0.05% of input. Custom
linearization user programmable at 21 points. Two selected functions
and custom linearization can operate on the input signal.
OUTPUT: Analog, Two wire 4 to 20mA
MINIMUM OUTPUT RANGE: None
OUTPUT RANGING ADJUSTMENTS:
Analog Zero:
100% of sensor range, non-interlacing
Analog full scale: Normal or reverse acting
OUTPUT RESOLUTION:
Analog, 2.1 uA:
ACCURACY:
Includes repeatability, hysteresis, load and ambient temperature effect
and uncertainty of factory NIST traceable calibration equipment.
Enhanced accuracy calibrations available from the factory.
OUTPUT ACCURACY:
+/- 0.04% of the millivolt or ohm equivalent reading, or the
accuracy from the table below, whichever is greater.
Input type
Accuracy
Milliamp
+/-0.005mA
Millivolt
+/-0.008mV
Volt
+/-0.003V
Resistance
+/-0.12ohm
TRANSMITTER REPEATABILITY:
One half of accuracy.