Do not lay connecting cables in the direct vicinity of cables carrying higher
voltage or used to switch inductive or capacitive loads.
ESD sensitive component. Electrostatic discharge may damage the PT. Take
ESD precautions.
6.3 Connection Assignments
7. Commissioning
7.1 Supply Voltage
Conduit / Leads
Red
Black
Signal -
White
Exci
Green
Excitation -
7.1 Supply voltage
7.2 Calibration
7.3 Zero adjustment
7.4 Operation
Please read the entire manual prior to installation and use. Maximum excitation is 28
Vdc. Rated excitation is +/- 12 Vdc. Recommended warmup is 15 minutes.
Explosion hazard!
Deviation of the supply voltage from the value given in the technical specifications, or
false polarity, candamage the pressure transmitter and cause malfunctions that can
pose a risk of explosion.