6.5.1 EMC / compliant connection
- Earth the machine section with the screw-in trunnion / mounting hole for the PT
in accordance with regulations. The PT must be connected to earth via the screw-in
trunnion / mounting hole.
- Connect the shield of the connecting cable on both sides, making sure it conducts
with full and continuous contact.
- When introducing the connecting cable into an EMC compliant switch cabinet, for
example, connect the shield correctely (cable gland, conducting, full contact,
continuous) to the conductive housing or route it via built-in cable connector that is
also connected to the conductive housing.
- Connect unused cable cores or free cable ends correctely to the cable shield on both
sides.
For order numbers of EMC compliant cable connectors required for connecting the PT,
see chapter 9, Accessories.
6.6 Connection assignments
Standard models MDT 4xx x-xxx-xx:
Equipment connector: 6-pin male,
Bendix PT02A-10-6P
Cable socket:
PT06A-10-6S(SR)
Fig. 11: 6-pin female connector
Top view solder side
The connector housing is connected conductively to the housing of the PT.