6.4.1 Thermo Insulation
When installing the TDA please consider that the stem needs sufficient insulation. The
insulation must cover the entire stem (see drawing). The thermo insulation is
necessary for the correct temperature measurement.
Fig. 10:
6.5 Electrical connection
Mounting and electrical connection of the PT must be
done by specialists with EMC training, following all applicable
regulations, and in
pressureless, voltage-free
condition with the
machine switched off.
The machine must be secured against being switched back on!
ESD sensitive component. Electrostatic discharge may damage the
PT. Take ESD precautions.
Avoid laying the power cable in the direct vicinity of cables carrying
higher voltages or switching inductive or capacitive loads.
An EMC compliant power supply must be used. The electrical
connection must comply with EMC requirements.
If the electrical connection is not made as described in chapter
6.5.1, or if cables / cable connectors / cable glands other than
those stipulated by
DYNISCO
are used,
DYNISCO
cannot guarantee
that EMC requirements will be satisfied.