Scheduled Maintenance
7-3
Maintenance and
Troubleshooting
onto the stop blocks. Loosen the bolts on the crosshead, shake the crosshead and
then re-tighten. If the crosshead still does not slide freely, contact Instron Dynatup.
•
Check and tighten the screws on the velocity detector and verify that the flag on the
crosshead passes through the center of the velocity detector slot.
•
Check the air filter/water trap mounted on the rear motor enclosure. Push up on the
valve located at the bottom of the bowl to drain.
•
Check the condition of all electrical wiring and conduits including the power cord
and pendent cable.
•
Check the air system for air leaks. Investigate any persistent hissing.
•
Check that the interlock switches are operational. Open each door to verify that the
yellow pendent light goes off. If not, check the interlock switch and immediately
suspend use of the machine until repairs are made.
•
Check the shock absorbers and/or stop blocks for damage. The shock absorbers
should not leak oil and should offer mild resistance when compressed by hand. The
elastomeric stop block tops should be intact. Replace the tops immediately when
damaged. Variances in the performance of the two stop block tops or shock
absorbers can damage the crosshead.
•
Visually check the tup, tup insert and tup cables for damage.
•
Check the safety enclosure door handles, hinges and the back safety enclosure
mounting bolts for any looseness. Tighten, if necessary.
•
Check that the two large main springs are securely attached to their connecting rods.
The springs are exposed when the air supply is turned off.
•
Check that the striker plates are securely held in place. These plates are mounted to
the bottom of the crosshead and protect the lower crosshead from the shock absorbers
or stop blocks.
•
Check the tightness of the guide columns. They should not be able to turn. If the
guide columns are loose, contact Instron Dynatup.
•
Release cable tension and check that the sheaves in the hoist pulley bracket rotate
freely. These pulleys can be reached through the opening in the back of the top
enclosure.
•
Tighten all bolts on fixtures, baseplate, tablebase, etc.
Содержание 8250
Страница 1: ... M14 13655 EN Operating Instructions Instron Dynatup 8250 Drop Weight Impact Tester ...
Страница 10: ...Preliminary Pages M14 13655 EN x ...
Страница 29: ...Connecting Utilities and Initial Set up 3 7 Installation Figure 3 3 Motor Enclosure ...
Страница 34: ...Initial Machine Operation and Checkout M14 13655 EN 3 12 ...
Страница 49: ...6 1 Operation Chapter 6 Chapter 6 Operation Outline Performing a Test 6 2 ...
Страница 54: ...Performing a Test M14 13655 EN 6 6 ...
Страница 58: ...Scheduled Maintenance M14 13655 EN 7 4 Figure 7 1 Lubrication Points LUBRICATION POINTS ...
Страница 62: ...Troubleshooting M14 13655 EN 7 8 ...
Страница 64: ...Glossary M14 13655 EN A 2 ...