TFX Error Codes
Revised 2-22-2002
Code Number
Description
Correction
Warnings
0001
Serial number not present
Hardware serial number has become inoperative – system
performance will not be influenced.
0010
Signal Strength is below Signal
Strength Cutoff entry
Low signal strength is typically caused by one of the following:
•
Empty
pipe
•
Improper programming/incorrect values
•
Improper transducer spacing
•
Non-homogeneous pipe wall
0011
Measured Speed of Sound in the liquid
is greater than 10% different than the
value entered during meter setup
Verify that the correct liquid was selected in the BASIC menu.
Verify that pipe size parameters are correct.
0020
Heat Flow Units of measure have been
selected and an RTD module has not
been installed
Verify that RTD Module PN D020-1045-106 has been installed
in one of the I/O meter slots. Verify that OUTPUT1 or OUTPUT
2 has been configured for RTD measurements.
Class C Errors
1001
System tables have changed
Initiate a meter RESET by cycling power or by selecting
SYSTEM RESET in the SEC MENU.
1002
System configuration has changed
Initiate a meter RESET by cycling power or by selecting
SYSTEM RESET in the SEC MENU.
Class B Errors
3001
Invalid hardware configuration
Upload corrected file
3002
Invalid system configuration
Upload corrected file
3003
Invalid strategy file
Upload corrected file
3004
Invalid calibration data
Recalibrate the system
3005
Invalid speed of sound calibration data
Upload new data
3006
Bad system tables
Upload new table data
3007
Data Logger is off or not present
If desired, insert data logger and configure within the Datalog
Operations Menu. If logger is not present, configure I/O port for
no logger.
3010
One or more channels are not
responding (Multi-channel meters only)
Display indicates which secondary units are not communicating
with Master meter. Verify wiring, configuration and address of
secondary instrument.
3011
All channels are not responding (Multi-
channel meters only)
Verify wiring, configuration and address of secondary
instruments.
Class A Errors
4001
Flash memory full
Return unit to factory for evaluation