XB-10/BIOmaster/VIS-20/ENVmaster
instruction manual
28
TIME scan parameter setting
DATA MODE
mode setting, including %T or ABS
SAMPLE NAME
sample name, less than 8 characters
WL(nm)
scanning wavelength
SCAN TIME(s)
scan time. (unit: s). more than 60s, max: 60000s
INTERVAL(s)
scan interval. AUTO, 0.5s, 1s, 2s, 5s, 10s, 100s. AUTO interval, when scan time is
less than 1000s, scan interval is 1s. When scan time is 1000-2000s, scan interval is
2s. When scan time is 5000-20000s, scan interval is 10s. When scan time is more
than 20000s, scan interval is 100s.
Y-MAX
ordinate ceiling of spectrogram
Y-MIN
ordinate low of spectrogram, the instrument will automatically adjust the small value
to the ordinate low position.
PATHLENGTH
cuvette optical length, it can print, but cannot regulate absorbance value
OK
finish parameter setting, back to measure interface automatically
Save
save parameter (SD card)