XB-10/BIOmaster/VIS-20/ENVmaster
instruction manual
25
Wavelength scan parameter setting
DATA MODE
mode setting, including %T or ABS
SAMPLE NAME
sample name, less than 8 characters
PEAK THRESHOL
sentenced peak threshold setting, the longitudinal axis is greater than the threshold
value, the peak is displayed
START WL(nm)
scan initiative wavelength
END WL(nm)
scan end wavelength. The large wavelength values will automatically adjust to the
wavelength end position
INTERVAL(nm)
scan interval. AUTO, 0.2nm, 0.5nm, 1nm, 2nm, 5nm. AUTO interval, when scan
range is less than 500nm and time interval is 0.5nm. When scan range is more
than 500nm and time interval is 1nm. Peak value judge, large wavelength interval
adapts to judging flat peak, small one adapts to judge keen-edged peak.
Y-MAX
ordinate ceiling of spectrogram
Y-MIN
ordinate low of spectrogram, the instrument will automatically adjust the small
value to the ordinate low position.
PATHLENGTH
cuvette optical length, it can print, but cannot regulate absorbance value
OK
finish parameter setting, back to measure interface automatically
Save
save parameter