XB-10/BIOmaster/VIS-20/ENVmaster
instruction manual
15
Operation environment
Power
Voltage: AC 100V
~
240V, Frequency :50/60Hz±1Hz, Capacity: more than 200W
Grounding line resistance of 10Ω or less is required
The input power mutations, without interference from other large electrical equipment
Operating
temperature
5~35 Deg C
In order to perform a measurement under the most stable condition, we recommend
that the instrument is used in an air conditioned room of 20 ~ 25 Deg C.
Storage temperature:
-
20
~
55 Deg C
Operating
humidity
Less than 85%.
Because it is optical equipment, in order to ensure it in good working condition, we
recommend that the instrument is used under 30%
~
70% humidity.
Storage humidity: less than 85%
。
Workbench
Width: more than 550mm, Depth: more than 400mm, Load bearing: more than 15kg,
horizontal and reliable workbench.
If there are other auxiliary devices, it is necessary to enlarge workbench area.
More than 200mm space at the both sides of the instrument, avoid being close to the
wall.
Height of the instrument: about 160mm. In order to easy operation, please choose
suitable workbench.
Connect with
other equipment
Connect with PC
(
additional optional software and computer
)
. Indirect specify type
thermal printer
(
options
)
and other accessories.
When main unit connects with above equipments, please swithch off the power of the
main unit and other equipments.
All grounding cords are in good condition and can be connected with the grounding
line of the main unit.