Revision: JUNE 2017
70
12. Calibration of the HPFM
Normally, the HPFM will not require frequent calibrations. A factory calibration is recommend every two years. Dynamax
recommends that the calibrations only be performed by one of our fully trained technicians at the factory, or by one of the
trained and fully qualified Dynamax distributors. During calibration all data is saved and supplied with each HPFM CD-Rom,
The regression, and the calibration graphics are supplied and the documentation folder of the HPFM CD-Rom, and printed
for your convenience. Below is the example of the calibration data and graphics.
Calibration Regression Data for HPFM-12-111
Tuesday, February 28, 2012
Calibration Factors:
NIST. Pressure Calibration against certificate number: C-1950
Date/Time 2/27/2012 12:07
Index Pressure (kPa)
PT1 (mV)
PT2 (mV)
0
10.02
0.21382
0.2405
1
52.95
0.82739
0.86345
2
90.08
1.3564
1.40098
3
130.64
1.9355
1.98954
4
171.4 2.51592
2.57892
5
210.24
3.06891
3.13935
6
250.64
3.6454
3.72142
7
292.85
4.24646
4.32799
8
331.09
4.79199
4.87795
9
370.01
5.3308
5.42023
10
411.15
5.93252
6.02543
11
453.81
6.53818
6.63373
12
490.65
7.0611
7.15943
13
529.81
7.61586
7.71731
14
569.9
8.1795
8.28496
Содержание HPFM3
Страница 49: ...Revision JUNE 2017 49 b Click on the STATIC OFFSET set the new value to zero and SAVE ...
Страница 73: ...Revision JUNE 2017 73 Red Range ...
Страница 74: ...Revision JUNE 2017 74 Yellow Range ...
Страница 75: ...Revision JUNE 2017 75 Grey Range ...
Страница 76: ...Revision JUNE 2017 76 Brown Range ...
Страница 77: ...Revision JUNE 2017 77 Zero Adjustment Measurement ...