Revision: JUNE 2017
25
23. At this time, or if any future need arises for a quick hardware test, you may run the InstaCal test.
Check on the “Test” tab and select the ANALOG option. A diagnostic screen will appear.
The voltages at zero pressure should be as shown, less than .004 V (<1 micro V) at maximum pressure (560 kPa)
applied, the voltage is about .008 V, see the calibration section for details on other pressure/ voltage comparisons.
A valid temperature in degrees C, should appear as well.
Now that you have properly installed the program and configured the system, you are ready to run HPFM.
24. Double check your Calibration table. This step is to ensure that correct calibration table is copied into your HPFM directory
25. Go to your CD ROM drive, open “HPFM.ini” from the root directory with Notepad.
26. Then, go to your HPFM installed directory at C:\Program Files\Dynamax\HPFM and open the same file
name “HPFM.ini” with Notepad.