19. Select tab USB-2404-UI, check the settings, and if required make the following changes:
• ADC Timing Mode: Best 60Hz Rejects
(
DO NOT
use any other timing modes in the list)
• Temperature units: Celsius
Once these changes have been made click
OK
and proceed to the next step.
20. While under CH0 settings make the following changes:
• Channel Mode: Bridge
• Bridge Type: Full Bridge
• Range:+/-62mv/v
Once these changes have been made click
OK
and proceed to the next step