TM-051 PN 35903 EC Series Models 38105, 38100, 38110
Rev 04/16/08
13
5. Remove the glass filter (if installed in the 2000-EC) by removing one retaining strap held by two (2) 8-32 x
3/8" button head screws with 3/32 hex wrench. If the optional glass filter is installed, loosen the two (2) 8-32 x
3/8" button head screws in the remaining strap so that the filter glass can slide out.
6. Reach into the reflector and GRASP the flat area of lamp, at either end.
CAUTION:
ALWAYS USE A SOFT, CLEAN RAG, CLEAN PAPER TOWEL OR GLOVES WHEN HANDLING THE
LAMP. SKIN OILS LEFT ON THE LAMP WILL BURN INTO THE QUARTZ, REDUCING OUTPUT INTENSITY.
IF THE LAMP IS INADVERTENTLY TOUCHED, CLEAN LAMP THOROUGHLY WITH A SOFT CLEAN RAG AND
ALCOHOL.
7. PUSH lightly on the lamp toward the socket on the opposite end of the lamp
so that the end being grasped
can be LIFTED clear of the socket as shown below.
8. INSTALL THE NEW LAMP by following steps 5, 6 and 7 in reverse order.
IMPORTANT
:
Install the lamp such that the seal dimple on the bulge of the glass is facing towards the reflector surface.
Avoid touching the quartz tube with your fingers.
NOTE:
LAMP HAS NO POLARITY
9. Record the serial number of the unit and the hour-meter reading in the Bulb History Record.
10. Replace the lamp/reflector assembly housing on the mounting stand by engaging the mounting bracket.
11. Secure the thumbscrew when the lamp/reflector assembly housing is at the proper height.
12. Reconnect the power cord to the appropriate power source.
13. Turn Power Switch to “ON” position and allow at least 5 minutes for lamp to warm up before using.