Dwyer Instruments GFM2 Series Скачать руководство пользователя страница 5

The  (+)  and  (-)  power  inputs  are  each  protected  by  a  300  mA  M  (medium  time-lag)
resettable fuse. If a shorting condition  or  polarity  reversal  occurs,  the  fuse  will  cut
power to the flow transducer circuit. Disconnect the power to the unit, remove the
faulty  condition,  and  reconnect  the  power.  The  fuse  will  reset  once  the  faulty
condition has been removed. DC Power cable length may not exceed 9.5 feet (3
meters). 

Use  of  the  GFM2  flow  transducer  in  a  manner  other  than  that  specified  in  this
manual  or  in  writing  from  Dwyer,  may  impair  the  protection  provided  by  the
equipment.

3. Principle of Operation

The stream of gas entering the Mass Flow transducer is split by shunting a small
portion of the flow through a capillary stainless steel sensor tube. The remainder of
the gas flows through the primary flow conduit. The geometry of the primary conduit
and the sensor tube are designed to ensure laminar flow in each branch.
According to principles of fluid dynamics the flow rates of a gas in the two laminar
flow conduits are proportional to one another. Therefore, the flow rates measured
in the sensor tube are directly proportional to the total flow through the transducer.

In order to sense the flow in the sensor tube, heat flux is introduced at two sections
of the sensor tube by means of precision wound heater-sensor coils. Heat is
transferred through the thin wall of the sensor tube to the gas flowing inside. As
gas flow takes place heat is carried by the gas stream from the upstream coil to
the downstream coil windings. The resultant temperature dependent resistance
differential is detected by the electronic control circuit. The measured temperature
gradient at the sensor windings is linearly proportional to the instantaneous rate
of flow taking place.

An output signal is generated that is a function of the amount of heat carried by
the gases to indicate mass-molecular based flow rates.

Additionally, the GFM2 Mass Flow Meter incorporates a precision analog
microcontroller and non-volatile memory that stores all hardware specific variables
and  up  to  10  different  calibration  tables.  The  flow  rate  can  be  displayed  in  23
different  volumetric  or  mass  flow  engineering  units.  Flow  meter  parameters  and
functions  can  be  programmed  remotely  via  the  RS-232/RS-485  interface  or
optional  Profibus  DP  interface.  GFM2  flow  meters  support  various  functions
including:  programmable  flow  totalizer,  low,  high  or  range  flow  alarm,  automatic
zero  adjustment  (activated  via  local  button  or  communication  interface),  2
programmable SPDT relays output, 0 to 5 VDC / 4 to 20 mA analog outputs (jumper
selectable), self diagnostic alarm, 36 internal and user defined K-factor. Optional
local  2x16  LCD  readout  with  adjustable  back  light  provides  flow  rate  and  total
volume  reading  in  currently  selected  engineering  units  and  diagnostic  events
indication.

SPECIFICATIONS
Service: 

Clean gases compatible with wetted parts.

Wetted Materials: 

GFM2-X-X-A: Anodized aluminum, brass, 316 SS fluoroelastomer O-rings;
GFM2-X-X-S: 316 SS, and fluoroelastomer O-rings; Buna-N, EPR and PTFE 

O-rings optional.

Accuracy: 

±1% FS.

Repeatability: 

±0.25% FS.

Response Time: 

2 seconds to within ±2% of actual flow.

Output Signal: 

Linear 0 to 5 VDC (3000 Ω min. load impedance) and 4 to 20 mA

(500 Ω max. loop resistance).

Max. Particulate Size: 

5 microns. 

Temperature Limits: 

32 to 122°F (0 to 50°C).

Power Supply: 

11 to 26 VDC. 

Process Connections: 

1/8˝ compression fitting for flow rates ≤ 10 L/min; 1/4˝ for ≤

50 L/min; 3/8˝ for ≤ 100 L/min. 

Display: 

2 x 16 character LCD.

Pressure Limits: 

500 psig (34.5 bar).

Leak Integrity: 

1 x 10-9 smL/sec of helium. 

Weight: 

1.05 lb (0.48 kg).   

PIN   GFM2 FUNCTION

1       Common, Signal Ground 

For Pin 2 (4 to 20 mA return)

2       0 to 5 VDC or 4 to 20 mA 

Flow Signal Output

3       Relay No. 2 - Normally Open 

Contact

4       Relay No. 2 - Common 

Contact

5       Common, Power Supply

(- DC power for 11 to 26 VDC)

6       Relay No. 1 - Common 

Contact

7       Plus Power Supply

(+ DC power for 11 to 26 Vdc)

8       RS485 (-) (Optional RS232 TX)
9       RS232 Signal GND (RS-485 

GND Optional)

10     Do not connect 

(Test/Maintenance terminal)

11     Relay No. 2 - Normally Closed 

Contact

12     Relay No. 1 - Normally Open 

Contact

13     Relay No. 1 - Normally Closed 

Contact

14     Do not connect 

(Test/Maintenance terminal)

15     RS485 (+) (Optional RS232

RX) Shield Chassis Ground

In general, "D" Connector numbering patterns are

standardized. There are, however, some connectors with

nonconforming patterns and the numbering sequence on your mating 
connector may or may not coincide with the numbering sequence shown in our
pin configuration table above. It is imperative that you match the appropriate wires
in accordance with the correct sequence regardless of then particular numbers
displayed on the mating connector.

NOTICE

Make sure power is OFF when connecting or disconnecting
any cables in the system.

NOTICE

Page 4

Содержание GFM2 Series

Страница 1: ...thin 2 of actual flow Output Signal Linear 0 to 5 VDC 3000 min load impedance and 4 to 20 mA 500 max loop resistance Relay Rating 1 A 24 VDC Max Particulate Size 5 microns Temperature Limits 32 to 122...

Страница 2: ...ass Flow Meter Installation and Operating Instructions Bulletin F GFM2 DWYER INSTRUMENTS INC Phone 219 879 8000 www dwyer inst com P O BOX 373 MICHIGAN CITY INDIANA 46360 U S A Fax 219 872 9057 e mail...

Страница 3: ...7 6 2 Flow Path Cleaning 7 6 2 1 Restrictor Flow Element RFE 7 6 2 2 GFM2 Models 7 7 Calibration Procedures 8 7 1 Flow Calibration 8 7 2 Gas Flow Calibration of GFM2 Mass Flow Meters 8 7 2 1 Connecti...

Страница 4: ...thorough leak test of the entire system All GFM2 s are checked prior to shipment for leakage within stated limits See specifications in this manual 2 2 Electrical Connections GFM2 is supplied with a 2...

Страница 5: ...nterface or optional Profibus DP interface GFM2 flow meters support various functions including programmable flow totalizer low high or range flow alarm automatic zero adjustment activated via local b...

Страница 6: ...peration of the meter above 110 of maximum calibrated flow may increase recovery time 5 3 GFM2 Parameters Settings 5 3 1 Engineering Units Settings The GFM2 Mass Flow Meter is capable of displaying fl...

Страница 7: ...nitrogen for up to 36 gases are stored in the GFM2 see APPENDIX II In addition provision is made for a user defined conversion factor Conversion factors may be applied to any of the ten gas calibrati...

Страница 8: ...to disassemble the sensor If blockage of the sensor tube is not alleviated by flushing through with cleaning fluids please return meter for servicing 6 2 1 Restrictor Flow Element RFE The Restrictor...

Страница 9: ...s the sensor Average reading must be in the range 120 10 counts If it is not perform Auto Zero procedure see section 5 3 10 Zero Calibration 7 2 3 Gas Linearization Table Adjustment Gas flow calibrati...

Страница 10: ...RS 232 SOFTWARE INTERFACE COMMANDS 8 1 General The standard GFM2 comes with an RS 485 interface For the optional RS 232 interface the start character and two hexadecimal characters for the address mu...

Страница 11: ...command can be used only when absolutely no flow thru the meter and no earlier then 6 minutes after power up It can take several minutes to complete Unit will not respond to other commands when this...

Страница 12: ...R read current totalizer volume W warm up delay S setting status D disable sets K 1 I internal K factor U user specified factor S status FS mL sec mL min mL hr L sec L min L hr m3 sec m3 min m3 hr f3...

Страница 13: ...w measurement LCD display reading does correspond to the correct flow range but 0 5 Vdc output signal does not change always the same reading or around zero LCD display reading and 0 to 5 Vdc output v...

Страница 14: ...nsor temperature is too high MCU temperature is too high overload Fatal Error EEPROM or Auto Zero error Solution Flush clean or disassemble to remove impediments or replace the filter screen see secti...

Страница 15: ...rent Gas Table Number 0 9 Current Units of Measure 0 22 Alarm Mode E Enabled D Disabled Low Flow Alarm Setting FS 0 Disabled High Flow Alarm Setting FS 0 Disabled Flow Alarm Action Delay 0 3600 sec 0...

Страница 16: ...Flow in PFS 0 6 70 0 FS A D Value from Sensor counts Actual Flow in PFS 0 7 80 0 FS A D Value from Sensor counts Actual Flow in PFS 0 8 90 0 FS A D Value from Sensor counts Actual Flow in PFS 0 9 100...

Страница 17: ...n 114 C2Cl2F4 Freon 115 C2ClF5 Freon C318 C4F8 Germane GeH4 Germanium Tetrachloride GeCl4 Helium He 50 L min Helium He 1 50 L min Helium He 2 10 50 L min Index 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16...

Страница 18: ...butyl Aluminum C4H9 AL Titanium Tetrachloride TiCl4 Trichloro Ethylene C2HCl3 Trimethylamine CH3 3N Tungsten Hexafluoride WF6 Uranium Hexafluoride UF6 Vinyl Bromide CH2CHBr Vinyl Chloride CH2CHCl Xeno...

Страница 19: ...APPENDIX IV Component Diagram Top Component Side Bottom Component Side Page 18...

Страница 20: ...s required The Series GFM3 is not field serviceable and should be returned if repair is needed Field repair should not be attempted and may void warranty See 1 3 WARRANTY RETURN Refer to Terms and Con...

Страница 21: ...wyer Instruments Inc Printed in U S A 5 13 FR RA 444052 00 DWYER INSTRUMENTS INC Phone 219 879 8000 www dwyer inst com P O BOX 373 MICHIGAN CITY INDIANA 46360 U S A Fax 219 872 9057 e mail info dwyer...

Отзывы: