Dwyer Instruments GFM2 Series Скачать руководство пользователя страница 13

UART Error Codes:

1 - Not Supported Command or Back Door is not enabled.
2 - Wrong # of Arguments.
3 - Address is Out of Range (MR or MW commands).
4 - Wrong # of the characters in the Argument.
5 - Attempt to Alter Write Protected Area in the EEPROM.
6 - Proper Command or Argument is not found.
7 - Wrong value of the Argument.
8 - Reserved.
9 - Manufacture specific info EE KEY (wrong key or key is disabled).

9. TROUBLESHOOTING
9.1 Common Conditions

Your GFM2 Digital Mass Flow Meter was thoroughly checked at numerous quality
control points during and after manufacturing and assembly operations. It was
calibrated according to your desired flow and pressure conditions for a given gas
or a mixture of gases.

It was carefully packed to prevent damage during shipment. Should you feel that
the instrument is not functioning properly, please check for the following common
conditions first:
Are all cables connected correctly? Are there any leaks in the installation? Is the
power supply correctly selected according to requirements? When several meters
are used a power supply with appropriate current rating should be selected.

Were the connector pinouts matched properly? When interchanging with other
manufacturers' equipment, cables and connectors must be carefully wired for
correct pin configurations. Is the pressure differential across the instrument
sufficient?

Command
Name

Maintenance
Timer
Full Scale

LCD Back
Light

Read
EEPROM
Memory
Write
EEPROM
Memory

Description

Hours since last time unit was
calibrated.
Returns the full scale rated flow in
L/min. (Note: This term is not
multiplied by the current K factor)

LCD Back Light control
(0-100.0%)
0 - off
100 - Maximum Intensity

Reads the value in the
specified memory location.

Writes the specified value to the
specified memory location. Use
Carefully, can cause unit to
malfunction. (Note: Some
addresses are write protected!)

No.

11

12

13

14

15

Command

C

E

B

MR

MW

Argument 1

R (read timer)
C (set timer to zero)

0 to 100%

No Argument
<current settings>

0000 to 999
(Table Index)

0000 to 999
(Table Index)

Argument 2

Value

Argument 3

Argument 4

Response

<Value> (in Hours)
CC
<Value> (in L/min)

B:<Counts>
where:
Counts (0 – 4095)
B:<Value>
where:
Value (0 – 100.0)
<value>

MW,XXX,<Value>
where:
XXX=Table Index

No.

1

2

3

4

5

6

7

8

Indication

No zero reading after 15 min. warm up time and
no flow condition.
Status LED indicator and LCD display remains
blank when unit is powered up. No response
when flow is introduced from analog outputs 0 to
5 VDC or 4 to 20 mA.

LCD display reading and/or analog output 0 to 5
VDC signal fluctuate in wide range during flow
measurement.
LCD display reading does correspond to the
correct flow range, but 0-5 Vdc output signal
does not change (always the same reading or
around zero).
LCD display reading and 0 to 5 Vdc output
voltage do correspond to the correct flow range,
but 4 to 20 mA output signal does not change
(always the same or reading around 4.0 mA).
Calibration is off (more than ±1.0% FS).

LCD display reading is above maximum flow
range and output volt age 0 to 5 VDC signal is
more than 5.0 VDC when gas flows through the
GFM2.
Gas flows through the GFM2, but LCD display
reading and the output voltage 0 to 5 Vdc signal
do not respond to flow.

Likely Reason

Embedded temperature has been
changed.
Power supply is bad or polarity is
reversed.

PC board is defective.
Output 0-5 Vdc signal (pins 2–1 of the D-
connector) is shorted on the GND or
overloaded.
Output 0 to 5 VDC schematic is burned
out or damaged.
Analog flow output scale and offset
variable are corrupted.
External loop is open or load resistance
more than 500 Ω.
Output 4 to 20 mA schematic is burned
out or damaged.
GFM2 has initial zero shift.

Sensor under swamping conditions (flow
is more than 10% above maximum flow
rate for particular GFM2).
PC board is defective.
The gas flow is too low for particular
model of GFM2.
GFM2 models: RFE is not connected
properly to the inlet fitting.
Sensor or PC board is defective.

Solution

Perform Auto Zero Procedure (see section 5.3.6 “Zero
Calibration”).
Measure voltage on pins 7 and 5 of the 15 pin D-connector. If
voltage is out of specified range, then replace power supply with a
new one. If polarity is reversed (reading is negative) make correct
connection.
Return GFM2 to factory for repair.
Check external connections to pin 2 – 1, of the D-connector. Make
sure the load resistance is more than 1000 Ω.

Return GFM2 to factory for repair.

Restore original EEPROM scale and offset variable or perform
analog output recalibration (see section 7.3).
Check external connections to pins 2 and 15 of the D-connector.
Make sure the loop resistance is less than 500 Ω.
Return GFM2 to factory for repair.

Shut off the flow of gas into the GFM2 (ensure gas source is
disconnected and no seepage or leak occurs into the meter). Wait
for 15 min. with no flow condition and perform Auto Zero calibration
Procedure (see section 5.3.7 “Zero Calibration”).
Lower the flow through GFM2 within calibrated range or shut down
the flow completely. The swamping condition will end automatically.

Return GFM2 to factory for repair.
Check maximum flow range on transducer’s front panel and make
required flow adjustment.
Unscrew the inlet compression fitting of the meter and reinstall
RFE (see section 6.2.2). NOTE: Calibration accuracy can be 
Return GFM2 to factory for repair.

9.2 Troubleshooting Guide

Page 12

Содержание GFM2 Series

Страница 1: ...thin 2 of actual flow Output Signal Linear 0 to 5 VDC 3000 min load impedance and 4 to 20 mA 500 max loop resistance Relay Rating 1 A 24 VDC Max Particulate Size 5 microns Temperature Limits 32 to 122...

Страница 2: ...ass Flow Meter Installation and Operating Instructions Bulletin F GFM2 DWYER INSTRUMENTS INC Phone 219 879 8000 www dwyer inst com P O BOX 373 MICHIGAN CITY INDIANA 46360 U S A Fax 219 872 9057 e mail...

Страница 3: ...7 6 2 Flow Path Cleaning 7 6 2 1 Restrictor Flow Element RFE 7 6 2 2 GFM2 Models 7 7 Calibration Procedures 8 7 1 Flow Calibration 8 7 2 Gas Flow Calibration of GFM2 Mass Flow Meters 8 7 2 1 Connecti...

Страница 4: ...thorough leak test of the entire system All GFM2 s are checked prior to shipment for leakage within stated limits See specifications in this manual 2 2 Electrical Connections GFM2 is supplied with a 2...

Страница 5: ...nterface or optional Profibus DP interface GFM2 flow meters support various functions including programmable flow totalizer low high or range flow alarm automatic zero adjustment activated via local b...

Страница 6: ...peration of the meter above 110 of maximum calibrated flow may increase recovery time 5 3 GFM2 Parameters Settings 5 3 1 Engineering Units Settings The GFM2 Mass Flow Meter is capable of displaying fl...

Страница 7: ...nitrogen for up to 36 gases are stored in the GFM2 see APPENDIX II In addition provision is made for a user defined conversion factor Conversion factors may be applied to any of the ten gas calibrati...

Страница 8: ...to disassemble the sensor If blockage of the sensor tube is not alleviated by flushing through with cleaning fluids please return meter for servicing 6 2 1 Restrictor Flow Element RFE The Restrictor...

Страница 9: ...s the sensor Average reading must be in the range 120 10 counts If it is not perform Auto Zero procedure see section 5 3 10 Zero Calibration 7 2 3 Gas Linearization Table Adjustment Gas flow calibrati...

Страница 10: ...RS 232 SOFTWARE INTERFACE COMMANDS 8 1 General The standard GFM2 comes with an RS 485 interface For the optional RS 232 interface the start character and two hexadecimal characters for the address mu...

Страница 11: ...command can be used only when absolutely no flow thru the meter and no earlier then 6 minutes after power up It can take several minutes to complete Unit will not respond to other commands when this...

Страница 12: ...R read current totalizer volume W warm up delay S setting status D disable sets K 1 I internal K factor U user specified factor S status FS mL sec mL min mL hr L sec L min L hr m3 sec m3 min m3 hr f3...

Страница 13: ...w measurement LCD display reading does correspond to the correct flow range but 0 5 Vdc output signal does not change always the same reading or around zero LCD display reading and 0 to 5 Vdc output v...

Страница 14: ...nsor temperature is too high MCU temperature is too high overload Fatal Error EEPROM or Auto Zero error Solution Flush clean or disassemble to remove impediments or replace the filter screen see secti...

Страница 15: ...rent Gas Table Number 0 9 Current Units of Measure 0 22 Alarm Mode E Enabled D Disabled Low Flow Alarm Setting FS 0 Disabled High Flow Alarm Setting FS 0 Disabled Flow Alarm Action Delay 0 3600 sec 0...

Страница 16: ...Flow in PFS 0 6 70 0 FS A D Value from Sensor counts Actual Flow in PFS 0 7 80 0 FS A D Value from Sensor counts Actual Flow in PFS 0 8 90 0 FS A D Value from Sensor counts Actual Flow in PFS 0 9 100...

Страница 17: ...n 114 C2Cl2F4 Freon 115 C2ClF5 Freon C318 C4F8 Germane GeH4 Germanium Tetrachloride GeCl4 Helium He 50 L min Helium He 1 50 L min Helium He 2 10 50 L min Index 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16...

Страница 18: ...butyl Aluminum C4H9 AL Titanium Tetrachloride TiCl4 Trichloro Ethylene C2HCl3 Trimethylamine CH3 3N Tungsten Hexafluoride WF6 Uranium Hexafluoride UF6 Vinyl Bromide CH2CHBr Vinyl Chloride CH2CHCl Xeno...

Страница 19: ...APPENDIX IV Component Diagram Top Component Side Bottom Component Side Page 18...

Страница 20: ...s required The Series GFM3 is not field serviceable and should be returned if repair is needed Field repair should not be attempted and may void warranty See 1 3 WARRANTY RETURN Refer to Terms and Con...

Страница 21: ...wyer Instruments Inc Printed in U S A 5 13 FR RA 444052 00 DWYER INSTRUMENTS INC Phone 219 879 8000 www dwyer inst com P O BOX 373 MICHIGAN CITY INDIANA 46360 U S A Fax 219 872 9057 e mail info dwyer...

Отзывы: