8 | Maintenance and servicing
OPM BWT PERMAQ® Pico Rev. 2.0 EN
EN
38
Preparing for disinfection
CAUTION
Risk of injury due to corrosive substances
Contact with corrosive substances can cause permanent damage to the eyes, skin and respir-
atory system.
ü
Only authorised and trained personnel is permitted to work with chemicals.
1. Before using chemicals, consult the relevant safety data sheets and take the necessary
safety measures.
2. Wear the recommended protective equipment.
3. Wash your hands thoroughly after working with chemicals.
4. In the case of contact with the eyes, rinse the eyes immediately with plenty of clear water
and consult a doctor.
To prepare for disinfection, proceed as follows:
1) Open the device door.
2) Switch off the device at the
main switch
.
ð
Wait for two to three minutes until the pressure is released.
3) Unscrew the filter housing with the filter key.
CAUTION! Filter housing is filled with water.
4) Empty and clean the filter housing.
5) Remove the old pre-filter and insert the new pre-filter into the filter housing.
6) Fill the quantity of AQUARIS DES stipulated in the
Table 6: Dosage of chemical dis-
infectant [
}
000] between the filter and the filter housing.
7) Insert the housing with the pre-filter into the filter head. Ensure that it is centred correctly
during the insertion.
8) Screw on the filter housing again by hand.
9) Close the device door.
10
)
Switch on the device at the
main switch
.
ð
The device is ready for disinfection.
Dosage of chemical disinfectant AQUARIS DES
Pico 10
6 tablets
Pico 20
9 tablets
Pico 30
9 tablets
Pico 40
12 tablets
Pico 50
15 tablets
Pico 60
18 tablets
Pico 70
20 tablets
Pico 80
22 tablets
Pico 90
24 tablets
Table
6: Dosage of chemical disinfectant
Performing the disinfection
To perform the disinfection, proceed as follows:
1) In the main menu, use the
↓
arrow
to select
DISINFECTION int.
and press the
OK but-
ton
to confirm.