4
9000-615-32/02 1808V002
Important information
2 Safety
Dürr Dental has designed and constructed this
unit so that when used properly and for the in-
tended purpose it does not pose any danger to
people or property Nevertheless, residual risks
can remain You should therefore observe the
following notes
2.1 Intended purpose
This device is a piezo-operated ultrasonic de-
vice for use in dental applications It is mainly
used for the treatment of periodontal defects In
addition, the device is used in the area of proph-
ylaxis, peri-implantitis treatment as well as den-
tal hygiene
Scaler handpiece applications
– Subgingival and supragingival removal of
dental calculus and concrement
The piezo-ceramic drive of the Vector Scaler al-
lows efficient removal of deposits while provid-
ing the best and gentlest possible protection of
sensitive tissue structures The ergonomic de-
sign of the handpiece offers six long-lasting high
performance LEDs to provide the best possible
illumination even into areas that are difficult to
see
2.2 Intended use
The ultrasonic device is designed for use in peri-
odontology, for the removal of plaque and for
the cleaning of tooth surfaces This is done via
cavitation, polishing, grinding and scraping Fur-
ther assistance for the treatment can be provid-
ed by using hydroxyl and/or fluorapatite as a
polishing agent in periodontology Only agents
approved by the manufacturer may be used
The treatment is gentle on the teeth and almost
painless
2.3 Improper use
WARNING
Risk of explosion due to ignition of
combustible materials
i
Do not operate the unit in any rooms
in which inflammable mixtures may be
present, eg in operating theatres
Any other usage or usage beyond this scope is
deemed to be improper The manufacturer ac-
cepts no liability for damages resulting from im-
proper use In these cases the user/operator will
bear the sole risk
Amplitude reduction
Amplitude increase
Switch off and de-energise the unit
(eg unplug from mains)
Wear protective gloves
Wear protective goggles
Use a face mask
Use protective clothing
1.2 Copyright information
All circuits, processes, names, software pro-
grams and units mentioned in this document are
protected by copyright
The Installation and Operating Instructions must
not be copied or reprinted, neither in full nor in
part, without written authorisation from Dürr
Dental
EN