8 Commissioning
8.1 Function check
To finish the initial start-up process, all of the
connections must be checked to make sure
they are securely seated and leak-tight
Check for correct operation:
– Operating panel
– Flexible foot switch
– Visual and acoustic signals
Performing pairing between the flexible foot
switch and the device:
i
Switch on the device
3 s
1.
2.
LIQUID
PEDAL
FUNCTION
CLEAN
RINSE
POWER
LIQUID
PEDAL
FUNCTION
i
LED PEDAL flashes in orange
i
Press the flexible foot switch for around 3 sec-
onds until the orange PEDAL LED goes out
Result:
After successful pairing the unit is immediately
ready for operation
If nothing happens when the flexible foot
switch is operated, it is possible that the
foot switch being used is paired with a
different device In this case cancel the
pairing and repeat the process
Cancelling a pairing
Requirements:
– No foot switch cable must be connected to
the device or to the flexible foot switch
– The orange PEDAL LED is not on or is not
flashing
i
Switch off the device
i
Touch the LIQUID operating panel, keep
touching it and switch on the device
Result:
When the orange PEDAL LED flashes, this
shows that the current pairing has now been
cancelled