9000-615-32/02 1808V002
21
Usage
Settings on operating panel
The settings are changed by touching but with-
out applying pressure
The LEDs for POWER and LIQUID are
only active and light up when a hand-
piece is inserted
If no handpiece is inserted, the settings
for POWER and LIQUID cannot be ad-
justed
ON/Standby
The unit is switched on and switched to standby
mode via the operating panel
ON / Standby
The unit is equipped with an automatic standby
mode in order to save electricity When the unit
is left unused for 30 minutes it switches off
automatically
POWER
The power is adjusted from 1 to 5; the settings
are displayed via the 5 LEDs:
LED dis-
play
Power in %
1
20
2
40
3
60
4
80
5
100
The power is set to level 5 as the default factory
setting on delivery
The adjustable operating frequency for the
Scaler handpiece is 20 - 120
m
m
LIQUID
The water amount can be selected from 3 set-
tings and is displayed via 3 LEDs:
LED dis-
play
Water amount in ml/min
1
approx 30
2
approx 37
3
approx 45
RINSE
After every treatment the system must be rinsed
with water
The rinsing process is started by touching the
RINSE button; it ends automatically after ap-
prox 30 seconds
If a rinsing process is running it can be interrupt-
ed at any time by touching the RINSE button
LED flashing: the vibration behaviour
of the instrument is impaired
– Contact pressure of the instru-
ment is too high during treatment
– reduce the contact pressure
– Check the instrument for signs of
wear or bending
18
LED in the fluid container
LED on: normal operation, fluid con-
tainer filled enough
LED flashing: Fluid level low
If the minimum fluid level is reached,
the LED of the fluid container will
start to flash and an acoustic warn-
ing signal will sound (3x high-
pitched sound)
Acoustic signals
Audible sig-
nals
Trigger/situation
Clicking noise
– Touching of the operating
panel
– Function activated,
eg RINSE, CLEAN
Long, low
audible signal
– Function cannot be
performed
Warning signal,
3x high-
pitched audible
signal
– Fluid level at minimum
– Waiting for fluid during the
cleaning process
10.2 Adjustment options
Operating panel
CLEAN
RINSE
POWER
LIQUID
PEDAL
FUNCTION