9000-615-32/02 1808V002
27
Usage
i
Touch the CLEAN button for at least 2 sec-
onds
CLEAN
1x
+
+
LED CLEAN and the LED in the fluid con-
tainer flash and a clicking sound is heard
The device is cleaned for approx 10 minutes
with
Vector cleaner
until the fluid container
is empty The process ends automatically
LED RINSE lights up and a cyclically repeat-
ing warning signal sounds as an indication
that the device must be rinsed with water af-
ter the cleaning with
Vector cleaner
+
...
RINSE
45 s
45 s
3x
3x
RINSE (rinsing/disinfecting):
As a result of the rinsing with water the
special cleaner
Vector cleaner
is re-
moved from the system Any residue of
the cleaning agent could cause irritation
in the patient
i
Top up the fluid container to the upper mark-
ing with water
12.2 Activating the cleaning process
of the device
We recommend that cleaning is per-
formed every 4 weeks The cleaning
process can be started at any time as re-
quired
After approx 30 operating the CLEAN LED will
light up on the operating panel to show that
cleaning is required
A cleaning cycle comprises two steps that need
to be started one after the other:
– CLEAN
– RINSE (rinsing/disinfecting)
The cleaning process is not finished until
both steps have been successfully per-
formed one after the other
CLEAN:
i
Pull off the hose connection from the hand-
piece
i
Place the handpiece hose in a sink or in a
suitable container
i
Pour 2 sealing caps (approx 40 ml) of
Vector
cleaner
undiluted into the empty fluid container
2x
EN