17. Troubleshooting
25
Characteristic
1.|
After insertion of the batteries
no acoustic signal sounds
2.|
During opening and closing of the
faceshield no signal sounds
3.|
The LED lights up if a hand is held
below the outlet and no media is
dispensed
4.|
The appliance seems to function
normally but no medium is
dispensed
5.|
An acostic signal onyl sounds if
the hand is held below the outlet
and no media is dispensed
6.|
After insertion of the locking plate
the appliance does not function
as usual
7.|
The appliance doses too much or
not enough medium
possible cause
|
Batteries are low.
|
Batteries were not inserted correctly.
|
Faceshield is closed while insertion.
|
Batteries are low.
|
No batteries inserted.
|
Battery contact interrupted.
|
Appliance defective.
|
Operating error.
|
Appliance recognized a disturbance
(unwanted radiation influance or
electric field).
|
No receptacle is inserted.
|
No pump is inserted.
|
Pump is clogged.
|
Receptacle is empty.
|
Pump lost contact to the medium.
|
An unsuitalbe medium has been used.
|
Batteries are low.
|
Appliance has not yet adjusted to the
locking plate.
|
Dosage adjustment is not adjusted
correctly.
solution
|
Insert new batteries.
|
Open the battrey box, insert batteries
according to the diagram. Pay attention
to polarity and the sequence.
|
Remove batteries, open faceshield,
insert new batteries again.
|
Insert new batteries.
|
Insert batteries.
|
Restore battery contact.
|
Call after-sales service.
|
Operate appliance as described above.
Pay attention to the sequence of
actuation in the sensor areas. Avoid
contact with the faceshield.
A)
Remove all electromagnetically charged
or emitting appliances out of the reach
of the sensor.
B)
Open and close the face shield
and quickly remove hands and / or
other disturbing objects.
Wait for a short time until the dis-
penser has gauged. Then actuate the
dispenser as usual.
C)
Remove receptacle and pump and
reinsert them.
D)
Make sure that pump sits correctly.
|
Insert full receptacle.
|
Insert
Touchless
pump into the dispenser.
|
Remove pump and clean it.
|
Replace receptacle.
|
Carry out the dosage process several
times. If it dones’nt function, the pump
is possibly clogged.
|
Use different medium.
Clean pump thoroughly before.
|
Replace batteries.
|
Wait (max for a few minutes).
|
Carry out dosage adjustment on the
pump.
EN