7. Going into operation

Open faceshield.

Open battery cover by releas-
ing the lower strap an the front 
side underneath the base 
angle. Open the batt. cover.

7. Going into operation

7.1

 Insert the batteries

The back of the dispesner must therefor be accessible (mounting plate removed).

Insert the batteries in the 
dispenser. 

Pay attention 

to correct polarity and 
sequencs as above!

1

2

4

3

5

 

•  If the dispesner is not used for a longer period than 3 months, batteries should be  

 

  removed in order to prevent unwanted discharge.

Unhinge the cover and put it 
beside. 

•   After correct insertion of the batteries an acoustic signal sounds.
•   Slide the dispenser from the above onto the mounting plate. 
•   Close the faceshield: now the dispenser starts with the first gauging.
•   The noise level of the unit is below 70 dB (A).

 

•  If the batteries have not been inserted with closed face shield, the measuring procedure has to 

 

  be carried out once again after having put the dispenser on the mounting plate by shortly  

 

  opening  and closing the faceshield such as with the change of the pump.  

 

  This procedure is explained below in more detail. 

7.2

 Use of the appliance

 • 

Prevent unwanted pump strokes by opening the face shield and keeping it open during  

 

  exchange. Body parts as well as objects have to be removed immediately for about     

 

  10 seconds after having closed the faceshield because dispenser adjusts itself.  

 

      

 

  The LED lights while the dispenser gauges. As soon as the light is out the dispenser is  

 

  ready for operation.

  

After change of receptacle or pump, the medium has to be primed with several pump  

 

  strokes (repeated pumping operation necessary). 

 

 

With the correct opening and closing of the faceshield, a (different) acoustic signal    

   sounds.

 

• 

After sliding dispenser onto mounting plate, insert pump and close the faceshield.     

 

  The dispenser will adjust to the pump stroke size and the environment.

 

• 

Insert a receptacle filled with medium of your choice into the dispenser. 

•  

Make sure that receptacle is in correct safe position and cannot fall out the dis 

 

   penser.

 

• Now you can use the appliance. 

EN

Содержание Touchless T 1000

Страница 1: ...EN 2014 07 Universal Dispenser Touchless T 400 T 1000 Installation and Operating instructions ...

Страница 2: ...e Non observance may cause damage to persons or defects to the device Notice Supplementary information about the operation of the unit The manufacturer does not assumy any liability for damages in the case of Defects at the appliance caused by mechanical influences or overvoltage Changings at the appliance without express approval of the manufacturer Use for the other purposes than that mentioned ...

Страница 3: ... Do not expose the appliance to thermal radiation and heat Only foll the appliance with the indicated substances in accordance with the intended use There are no dismountable parts wioth the exception of pump receptacle battery cover mounting plate batteries locking plate accessories The appliance is not autoclavable and cannot be cleaned in the dishwasher The apparatus must not be cleaned with wa...

Страница 4: ...e it is used The dispenser only meters with closed faceshield The emergency lever has to be actuated exclusively in case of a breakdown of automatic function 4 Dispenser components scope of delivery EN Model No Universal Dispenser Touchless T 400 6000 049 27 Universal Dispenser Touchless T 1000 6000 049 28 4 1 Asseccories Accessory for Touchless Drip tray 6000 049 29 4 2 Filling goods for the Univ...

Страница 5: ... on the floor Prior to use of new filling media tests are required Please pay attention to that during innstallation no radiation emittent object i e X ray apparatus mobile telephone ultrasound device direct illumination over the dispenser etc is in the near proximity Electromagnetic interference sources and or radiating electric fields may impair the function Prior to the mounting of the dispense...

Страница 6: ...opening and press down mounting tongue of the mounting plate Pull dispenser upwards with pressed mounting tongue 6 2 Mounting For the mounting and going into operation of the Universal Dispenser Touchless please keep the following steps in sequence Fix mounting plate at the wall in desired position by means of dowels and crews Drilling diagram see technical data Slide the dispenser from above onto...

Страница 7: ...r having put the dispenser on the mounting plate by shortly opening and closing the faceshield such as with the change of the pump This procedure is explained below in more detail 7 2 Use of the appliance Prevent unwanted pump strokes by opening the face shield and keeping it open during exchange Body parts as well as objects have to be removed immediately for about 10 seconds after having closed ...

Страница 8: ...ly when having your other hand first under the outlet If you approach with only one hand from above without holding the other had below the outlet the appli ance does not pump The actuation of the dispenser is effected if a hand appproaches from below under the outlet At a distance of approx 4 cm the dispenser releases automati cally a pumping process 9 Pump change The pump must only be changed if...

Страница 9: ... from the dispenser Insert the filled up receptacle from below into the dispenser Please pay attention to the correct position of the receptacle If necessary remove the locking plate before Please take care that only receptacle with an opening width of max 45mm is used Make sure that the bottle adapter sits tight onto the opening so that no gas can exhaust from the receptacle In order to avoid uni...

Страница 10: ...electromagnetic interference sources and false handling are recognized and indicated by the lighting up of the LED As long as the LED lights up the automatic pumping action is suppressed The LED also lights up during gauging When it expires the dispenser is ready for operation EN ...

Страница 11: ...exposed to water jets The appliance can not be cleaned in a washer disinfector and is not dishwasher proof and is not autoclavable The Touchless pump can be cleaned in a washer disinfector and is dishwasher proof and autoclavable However it must rest for 6 hours after such a treatment before it is reinserted in the dispenser 13 2 Exchange of battery Test the battery level First of all approach one...

Страница 12: ...g centre or ask your specialized dealer about applicable disposal regulations 16 Technical data Shroud aluminium anodized Lever stainless steel Mounting plate plastic Drip tray plastic stainless steel Pump plastic and stainless steel removable Wall mounting with dowels and screws Conformity RoHS compliant according to machinery directive 2006 42 EC EMC directive 2004 108 EC Manufacturer Ophardt Ir...

Страница 13: ...ance has not yet adjusted to the locking plate Dosage adjustment is not adjusted correctly solution Insert new batteries Open the battrey box insert batteries according to the diagram Pay attention to polarity and the sequence Remove batteries open faceshield insert new batteries again Insert new batteries Insert batteries Restore battery contact Call after sales service Operate appliance as descr...

Страница 14: ......

Страница 15: ......

Страница 16: ...DÜRR DENTAL AG Höpfigheimer Strasse 17 74321 Bietigheim Bissingen Germany Fon 49 7142 705 0 www duerr de info duerr de ...

Отзывы: