Configuration, installation, commissioning
6.4.2 Commissioning
First plug in the connection cable to the measuring probe, using the plug
connector and secure it with the cap nut.
The cap nut screwed connection is sealed with a rubber ring. The nut must be
well tightened
against the resistance of this rubber ring, to make sure of a
reliable seal at the connection.
The supply unit is now supplied with current, so that the purge air hose pushes
clean ambient air through the dust measuring lance. A flow of air must be
perceptible at the diaphragm openings.
Figure 6.12: Diaphragm openings
If this is the case, the dust measuring probe can be inserted into the duct
flange together with the flange seal. The red dot at the top of the welded-in
flange must lie alongside the red dot on the probe flange. The probe is then
fastened to the duct flange with four self-locking M10 nuts.
Before the connection cable can be disconnected and reconnected, the
miniature fuse holder for the fuse L1 in the supply unit must be swung open.
The measuring lance then performs a reset and an auto-initialisation. Any
measurement results collected previously but not saved are lost when this is
done!
Note also that removing the will stop the purge air supply! Therefore first take
the D
−
R 800 out of the duct that is being measured.
Figure 6.13: Miniature fuse holder
62
D
−
R 800
Содержание D-R 800
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 16: ......
Страница 24: ......
Страница 28: ......
Страница 30: ......
Страница 40: ......
Страница 46: ......
Страница 64: ......
Страница 66: ......
Страница 84: ...Modbus 84 D R 800 ...
Страница 94: ......
Страница 96: ......
Страница 130: ......
Страница 136: ... Technical data 136 D R 800 ...
Страница 137: ... Menu structure D R 800 137 11 Menu structure Figure 11 1 Menu structure D R 800 part 1 3 ...
Страница 138: ... Menu structure Figure 11 2 Menu structure D R 800 part 2 3 138 D R 800 ...
Страница 139: ... Menu structure D R 800 139 Figure 11 3 Menu structure D R 800 part 3 3 ...
Страница 140: ... Menu structure 140 D R 800 ...
Страница 142: ......
Страница 144: ...Declaration of Conformityand QAL1 Figure 13 2 QAL1 to DIN EN 14181 and DIN EN 14956 page 1 2 144 D R 800 ...
Страница 145: ...Declaration of Conformityand QAL1 D R 800 145 Figure 13 3 QAL1 to DIN EN 14181 and DIN EN 14956 page 2 2 ...
Страница 148: ...Glossary 148 D R 800 ...
Страница 154: ...GmbH Kollaustraße 105 22453 Hamburg www durag de ...