WARNING
Use of cables other than those approved by Duniway Stockroom
Corp. could lead to operation problems or safety hazards.
Figure 2.
Terranova® 752A back view
To properly connect the Terranova® 752A to the ion pump(s):
1.
Secure the ion pump cable end to the ion pump high voltage feedthrough(s)
2.
Secure the ion pump cable to the respective HIGH VOLTAGE OUT connector
3.
Secure the required secondary ground sense lead(s) to both the ion pump body
and respective GROUND SENSE stud (see
Safety Interlock Feature
)
WARNING
NEVER apply power to the Terranova® 752A until proper
grounding has been established.
User should always ensure the ion pump and high voltage cable are properly connected and
secured before use.