![DUNIWAY Stockroom Terranova 752A Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/duniway-stockroom/terranova-752a/terranova-752a_instruction-manual_2539777030.webp)
TELEPHONE:
650-969-8811
TOLL-FREE (US/Canada):
800-446-8811
FAX:
650-965-0764
EMAIL:
WEB:
30
of
38
Figure 8
. HV diode and jumper J2 positive high voltage position
8.
Remove the HV diode
9.
Locate jumper J2 of the respective channel. As shown in
Figure 8
, jumper J2
–
identified
by the dashed box
–
is next to the HV board heat sink.
10.
Remove jumper J2
11.
Re-install jumper J2 on the proper pin connector
Positive polarity:
J2 POS
Negative polarity:
J2 NEG
12.
Re-install the HV diode in the proper orientation
CHANNEL 1
Positive polarity:
Stripe should face towards the inside of the unit
Negative polarity:
Stripe should face towards the outside of the unit
CHANNEL 2
Positive polarity:
Stripe should face towards the outside of the unit
Negative polarity:
Stripe should face towards the inside of the unit
Failure to install the HV diode and J2 jumper to the same polarity will
result in damage to the high voltage transformer.
13.
Locate DIP Switch S1-1 for CHANNEL 1 or DIP Switch S3-8 for CHANNEL 2
–
see
Figure 3.