IQ Explorer 3
QUICK START GUIDE
10
SE
CTION 3 - OPERA
TION
STEP 3 -
Set the Cycle Initiation Mode
Click on the Process Settings tab. Select Manual from the pull down menu as shown in Figure 3-3.
STEP 4 -
Specify the Trigger Settings
Select Force from the TRIGGER type drop down box. Then, program the desired amount of force the press is to apply
before ultrasound is activated. Next, enter the Max Trigger Time. This provides a measure of safety insuring the welder
will terminate if the welder never achieves the programmed trigger force as shown in Figure 3-4.
Figure 3-3
Initiate Mode
Figure 3-4
Trigger Force Settings
NOTE:
This Initiate Mode selection assumes there is a stand-alone, benchtop system with
manual operate switches.
Process Settings tab