MAINTENANCE OF THE COMPRESSOR
The compressor does not need any preventive maintenance.
CLEANING OF THE JET NEBULIZING KIT
After each treatment, clean the jets by making the nebulizing kit operate with water. Then disconnect the
tubes coming from the compressor and clean the nebulizing kit with hot water.
The nebulizing kit can be cleaned in the dishwasher.
DISINFECTION OF THE NEBULIZING KIT
The dismantling and disinfection instructions are mentioned on the case-box.
The complete nebulizing kit, mask and tubes, can be disinfected in boiling water (20 minutes).
As a hygiene measure, a new complete nebulizing kit is sold at each prescription.
Nota
: after cleaning in boiling water the tubes colour can change (whiter) but this would not change the
good operating of your nebulizing kit.
7.
FAILURES REMEDIES AND WARRANTY
1
st
INCIDENT:
NO OPERATING NOISE CAN BE HEARD
ADVICE: check that
The unit is well connected to the mains supply,
The electrical switch is on the operating position,
The power delivered by the mains supply fits in with that needed by the unit (volts-Hz~),
The electrical current arrives to the socket (try with another electric device).
2
nd
INCIDENT:
NO AEROSOL COMES OUT OF THE NEBULIZING KIT
ADVICE: check that
The tank contains enough liquid depending of the jet nebulizing kit: the minimum of medicine is
indicated on the instructions for use of each nebulizing kit,
The tubes are well connected to the compressed air inlet of the nebulizing kit and to the compressed
air outlet of the compressor,
The additional air inlet (ventury hole) of the nebulizing kit is not closed,
The tubes are not pinched or blocked,
The jets are well cleared,
The jet nozzle is completely inserted,
Glass fragments did not fall into the nebulizer, in case of use of medicine glass bulbs.
Contact the manufacturer:
- if you need help to set up, use or maintain the medical device,
- to report any unexpected operation or event.
IN CASE OF OTHER PROBLEM, REFER TO YOUR SUPPLIER