11/12

 

MAINTENANCE OF THE COMPRESSOR 

The compressor does not need any preventive maintenance.  

 
 

CLEANING OF THE JET NEBULIZING KIT  

After each treatment, clean the jets by making the nebulizing kit operate with water. Then disconnect the 
tubes coming from the compressor and clean the nebulizing kit with hot water.  
The nebulizing kit can be cleaned in the dishwasher.  

 
 

DISINFECTION OF THE NEBULIZING KIT  

The dismantling and disinfection instructions are mentioned on the case-box. 
The complete nebulizing kit, mask and tubes, can be disinfected in boiling water (20 minutes).  
As a hygiene measure, a new complete nebulizing kit is sold at each prescription. 

 

Nota

: after cleaning in boiling water the tubes colour can change (whiter) but this would not change the 

good operating of your nebulizing kit.  
 

 

7. 

FAILURES REMEDIES AND WARRANTY 

 

1

st

 INCIDENT: 

NO OPERATING NOISE CAN BE HEARD  

ADVICE: check that  

 

The unit is well connected to the mains supply,  

 

The electrical switch is on the operating position,  

 

The power delivered by the mains supply fits in with that needed by the unit (volts-Hz~),  

 

The electrical current arrives to the socket (try with another electric device).  

 

2

nd

 INCIDENT:

 NO AEROSOL COMES OUT OF THE NEBULIZING KIT  

ADVICE: check that  

 

The  tank  contains  enough  liquid  depending  of  the  jet  nebulizing  kit:  the  minimum  of  medicine  is 
indicated on the instructions for use of each nebulizing kit, 

 

The tubes are well connected to the compressed air inlet of the nebulizing kit and to the compressed 
air outlet of the compressor, 

 

The additional air inlet (ventury hole) of the nebulizing kit is not closed, 

 

The tubes are not pinched or blocked,  

 

The jets are well cleared,  

 

The jet nozzle is completely inserted, 

 

Glass fragments did not fall into the nebulizer, in case of use of medicine glass bulbs. 

 

Contact the manufacturer: 
- if you need help to set up, use or maintain the medical device, 
- to report any unexpected operation or event.

 

 

IN CASE OF OTHER PROBLEM, REFER TO YOUR SUPPLIER  

 

Содержание PURENEB AEROSONIC: PURENEB AEROSONIC+

Страница 1: ...a CLASSIC OR STANDARD MODELS FOR LUNG OR BRONCHIAL TREATMENTS AEROBOX ref ABOX 0459 1998 AEROBOX ref ABOX 0459 2001 SONIC MODELS FOR ENT SINUSITIS TREATMENTS AEROSONIC ref AOHBOX 0459 2006 AEROSONIC r...

Страница 2: ...ion 2 4 Symbols used on devices 3 PRECAUTIONARY MEASURES 4 DESCRIPTION AND PERFORMANCES 4 1 AEROBOX and AEROBOX models 4 2 AEROSONIC and AEROSONIC models 4 3 Lists of accessories and spare parts 4 4 P...

Страница 3: ...0 C Room temperature T Operating 5 C T 40 C Relative humidity H Transport storage and operating 15 H 93 Atmospheric pressure P Transport storage and operating 700 hPa P 1060 hPa Essential performances...

Страница 4: ...kV 15 kV Electrical fast transient burst IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines 2 kV 1 kV Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to l...

Страница 5: ...n aerosol stream subjected to sound vibrations around 100 Hz is called sonic aerosol The fineness of the particles is unchanged Aerosol particles are submitted to 100 Hz sound vibrations in order to m...

Страница 6: ...upply make sure that the nebulizing kit is connected to the unit in order to avoid an abnormal noise If the compressor with the nebulizing kit is noisy do not hesitate to contact your supplier Do not...

Страница 7: ...e disconnect the device from the mains Make sure the device is not exposed to direct sunlight or excessive heat Do not touch the device for more than 10 seconds during use Do not touch the tube for mo...

Страница 8: ...nd AEROSONIC with automatic control EXTRANEB SONIC nasal NL11SN GB nebulizing kit with nasal piece mask Switch with sound vibrations or without stop Carrying bag on request ref MBOXAS 4 3 List of acce...

Страница 9: ...etry the granulometric distribution depends on the nebulizing system compressor nebulizing kit Examples CLASSIC ATOMISOR for broncho pulmonary treatment PERFORMANCES EXTRANEB NL9M NEBULIZING KIT WITH...

Страница 10: ...l switch on sonic aerosol position Sound vibrations automatically alternate with periods without sound vibrations AFTER AN AEROSOLS TREATMENT Switch off the compressor switch on O Disconnect both tube...

Страница 11: ...k that The unit is well connected to the mains supply The electrical switch is on the operating position The power delivered by the mains supply fits in with that needed by the unit volts Hz The elect...

Страница 12: ...ease contact your distributor A technical description is available on request Regarding BOX type devices keep the specific packaging and foam set for sending back the device 8 WASTE DISPOSAL Nebulizin...

Отзывы: