background image

 

 

3/12

 

 

1.

CLASSIFICATION and ELECTROMAGNETIC COMPATIBILITY 

 

Device class 

IIa 

Device class in electrical safety 

II 

Type of applied parts 

BF 

Protection level  

IP 30 
3: protected against ingress of solid foreign objects larger than 
2.5 mm; 
0: not protected against vertical water drops falls. 

Expected shelf life of accessories 

1 year 

Expected shelf life of the device 

5 years 

Weight  ABOX / ABOX+ 
 

AOHBOX /  

3,8 kg 
4,5 kg 

Dimensions 

L360xl250xh180 mm 

Voltage 

230 V; 50 Hz (ABOX can be available in 220V-50Hz, 220V-60Hz 
or 240V-50Hz) 

Consumption 

ABOX 

 

ABOX+ 
AOHBOX /  

85 VA 
105 VA 
110 VA 

Using pressure  ABOX / AOHBOX 

ABOX+ /  

> 400 hPa 
> 650 hPa 

Sound level 

< 70 dBa 

Flow rate  ABOX / AOHBOX 

ABOX+ 
 

5.15 L/min 
6.85 L/min 
6.50 L/min 

Running mode 

Intermittent (30 min ON / 30 min OFF) 

Room temperature (T) - Transport and storage 

-25°C < T < +70°C 

Room temperature (T) - Operating 

+5°C < T < +40°C 

Relative humidity (H) - Transport, storage and operating 

15% < H < 93% 

Atmospheric pressure (P) - Transport, storage and operating 

700 hPa < P < 1060 hPa 

 
Essential performances: provide pressure allowing to transform the medicine into aerosol. 
 
Applied part: mouth piece, nasal piece or mask 
 

Note:

 

The CE certificate can be supplied on request.  

 
 

ELECTROMAGNETIC COMPATIBILITY 

 

Electro medical devices require special precautions against EMC and must be installed and set to operate 
in accordance with the EMC information provided in this manual. 
WARNING: Avoid using this unit near or stacked onto other equipment as this may cause a malfunction. If 
this is necessary, this appliance and the other appliances should be monitored for normal operation. 
WARNING:  Use  of  accessories,  transducers  and  cables  other  than  those  specified  or  supplied  by  the 
manufacturer  of  this  device  may  cause  an  increase  in  electromagnetic  emissions  or  a  decrease  in  the 
immunity of this device and cause improper operation. 

Содержание PURENEB AEROSONIC: PURENEB AEROSONIC+

Страница 1: ...a CLASSIC OR STANDARD MODELS FOR LUNG OR BRONCHIAL TREATMENTS AEROBOX ref ABOX 0459 1998 AEROBOX ref ABOX 0459 2001 SONIC MODELS FOR ENT SINUSITIS TREATMENTS AEROSONIC ref AOHBOX 0459 2006 AEROSONIC r...

Страница 2: ...ion 2 4 Symbols used on devices 3 PRECAUTIONARY MEASURES 4 DESCRIPTION AND PERFORMANCES 4 1 AEROBOX and AEROBOX models 4 2 AEROSONIC and AEROSONIC models 4 3 Lists of accessories and spare parts 4 4 P...

Страница 3: ...0 C Room temperature T Operating 5 C T 40 C Relative humidity H Transport storage and operating 15 H 93 Atmospheric pressure P Transport storage and operating 700 hPa P 1060 hPa Essential performances...

Страница 4: ...kV 15 kV Electrical fast transient burst IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines 2 kV 1 kV Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to l...

Страница 5: ...n aerosol stream subjected to sound vibrations around 100 Hz is called sonic aerosol The fineness of the particles is unchanged Aerosol particles are submitted to 100 Hz sound vibrations in order to m...

Страница 6: ...upply make sure that the nebulizing kit is connected to the unit in order to avoid an abnormal noise If the compressor with the nebulizing kit is noisy do not hesitate to contact your supplier Do not...

Страница 7: ...e disconnect the device from the mains Make sure the device is not exposed to direct sunlight or excessive heat Do not touch the device for more than 10 seconds during use Do not touch the tube for mo...

Страница 8: ...nd AEROSONIC with automatic control EXTRANEB SONIC nasal NL11SN GB nebulizing kit with nasal piece mask Switch with sound vibrations or without stop Carrying bag on request ref MBOXAS 4 3 List of acce...

Страница 9: ...etry the granulometric distribution depends on the nebulizing system compressor nebulizing kit Examples CLASSIC ATOMISOR for broncho pulmonary treatment PERFORMANCES EXTRANEB NL9M NEBULIZING KIT WITH...

Страница 10: ...l switch on sonic aerosol position Sound vibrations automatically alternate with periods without sound vibrations AFTER AN AEROSOLS TREATMENT Switch off the compressor switch on O Disconnect both tube...

Страница 11: ...k that The unit is well connected to the mains supply The electrical switch is on the operating position The power delivered by the mains supply fits in with that needed by the unit volts Hz The elect...

Страница 12: ...ease contact your distributor A technical description is available on request Regarding BOX type devices keep the specific packaging and foam set for sending back the device 8 WASTE DISPOSAL Nebulizin...

Отзывы: