DT Research DT313 series Скачать руководство пользователя страница 18

Version: 1.0

 

 

 

Radiated RF 

IEC 61000-4-3 

 

 

10 V/m 

80 MHz to 2.7 GHz 

 

 

10 V/m 

 
 
 
d=1.2√P 80MHz to 800MHz  
d=2.3√P 800MHz to 2.7GHz  

Where 

P

 is the maximum output power rating of the transmitter 

in watts (W) according to the transmitter manufacturer and 

d

 is 

the recommended separation distance in metres (m). 

Field strengths from fixed RF transmitters, as determined by an 
electromagnetic site survey,

a

 should be less than the compliance 

level in each frequency range.

b

 

Interference may occur in the vicinity of equipment marked with 
the following symbol: 

 

NOTE 1       At 80 MHz and 800 MHz, the higher frequency range applies. 

NOTE 2       These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection 
from structures, objects and people. 

 
a      Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land 

mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically 

with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site 

survey should be considered. If the measured field strength in the location in which the  device is used exceeds the applicable 

RF compliance level above, the  device should be observed to verify normal operation. If abnormal performance is observed, 

additional measures may be necessary, such as re-orienting or relocating the  device. 

b     Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. 

 
 

Recommended separation distances between  

portable and mobile RF communications equipment and the device 

The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the 
user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF 
communications equipment (transmitters) and the device as recommended below, according to the maximum output power of the 
communications equipment. 

Rated maximum output power 

of transmitter 

(W) 

Separation distance according to frequency of transmitter 

(m) 

150 KHz to 80 MHz 

 

 

 

80 MHz to 800 MHz 

 

 

800 MHz to 2.7 GHz 

 

 

0.01 

0.12 

0.12 

0.23 

0.1 

0.38 

0.38 

0.73 

1.2 

1.2 

2.3 

10 

3.8 

3.8 

7.3 

100 

12 

12 

23 

For transmitters rated at a maximum output power not listed above, the recommended separation distance din metres (m) can be       
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the                  
transmitter in watts (W) according to the transmitter manufacturer. 
 
NOTE 1   At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 
NOTE 2   These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 

reflection from structures, objects and people. 

Содержание DT313 series

Страница 1: ...Version 1 0 Medical Cart Computer DT313 SERIES User s Manual...

Страница 2: ...e DT313 series is ready for use out of the box in its default configuration when powered by the power source provided The following discussion offers guidance on the hardware elements and features of...

Страница 3: ...Version 1 0 Customize Mobility Center Security Keyboard IEC 60417 5032 AC Atmospheric pressure limitation Humidity limitation Temperature limitation...

Страница 4: ...Transportation 20 C 60 C Operation 0 C 40 C Storage 20 C 60 C Humidity Transportation 0 90 Operation 0 90 Storage 0 90 Atmospheric Pressure Transportation 70 0 kPa to 106 0 kPa Operation 70 0 kPa to 1...

Страница 5: ...CC Rules Operation is subject to the following two conditions 1 This device may not cause harmful interference and 2 this device must accept any interference received including interference that may c...

Страница 6: ...intended use for general operator Not intended for patient used Attention Les Panel PC sont utilisation pr vue pour l op rateur Ne pas utiliser patient Grounding reliability can only be achieved when...

Страница 7: ...Version 1 0 Function Instructions 1 Input Output Ports and Button Functions...

Страница 8: ...NOT apply excessive pressure to the display screen We recommend using the Digital Pen optional to keep the screen clean Avoid prolonged exposure of the display panel to any strong heat source Whereve...

Страница 9: ...ovided in the software operating system Wireless Networking Wireless LAN The DT313SC MD DT313T MD is often delivered with an embedded user inaccessible 802 11ac WLAN adapter equipped with a hidden cus...

Страница 10: ...ntable cradle options These securable and mountable cradles provide an interface to VESA mounts and arms and to ports that may include USB Ethernet and pass through power For More Support Users can do...

Страница 11: ...igh temperature it will speed up chemical metabolism and shorten the battery service life Please don t store the battery in a temperature over 60 C It may cause the battery explode 4 Caution 1 The bat...

Страница 12: ...acement des piles au lithium ou les piles combustible par du personnel mal form Battery change 1 Remove the battery Push the switch up to unlock the latch then slide the latch left to remove the batte...

Страница 13: ...SB ports Instructions Connect the AC DC adapter cable to the DC input jack and then insert the Tablet system properly for charging Package Contents 1 Desktop Cradle 2 Basic Operation Guide Precautions...

Страница 14: ...20 C 60 C Humidity 0 90 non condensing The AC DC Adapter is not included in the standard package Specificati uons subject to change without notice Tablet Mont Cradle BASIC OPERATION GUIDE INTRODUCTION...

Страница 15: ...o the cradle first with the connector parts correctly lined up 3 When the bottom of the Tablet is in place push the Tablet against the back and into the Cradle 4 Push down on the Top Clasp so the top...

Страница 16: ...should be at Guidance and manufacture s declaration electromagnetic emission The device is intended for use in the electromagnetic environment specified below The customer of the user of the device s...

Страница 17: ...ued operation during power mains interruptions it is recommended that the device be powered from an uninterruptible power supply or a battery Power frequency 50Hz 60Hz magnetic field IEC 61000 4 8 30...

Страница 18: ...rformance is observed additional measures may be necessary such as re orienting or relocating the device b Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m Recommen...

Страница 19: ...hould be installed and operated with minimum distance 0cm between the radiator and your body This transmitter must not be co located or operating in conjunction with any other antenna or transmitter T...

Страница 20: ...AR la plus lev e signal e en vertu de cette norme lors de la certification de produit utiliser lorsqu il est correctement port sur le corps Le SAR est r alis sur une plaque mobile dt313 quip e d un mo...