![DT Research DT313 series Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/dt-research/dt313-series/dt313-series_user-manual_2535698016.webp)
Version: 1.0
I/O Port
Ethernet
RJ45 connector for Ethernet x 1
COM
1
USB 2.0
4
HDMI-out
1
DC-in
1
EMC Information
Guidance and manufacture’s declaration – electromagnetic immunity
The device
is intended for use in the electromagnetic environment specified below. The customer or the user of device
should
assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
2,4,6,8 kV
contact
2,4,8,15 kV
air
2,4,6,8 kV
contact
2,4,8,15 kV
air
Floors should be wood, concrete
or ceramic tile. If floor are
covered with synthetic material,
the relative humidity should be at
Guidance and manufacture’s declaration – electromagnetic emission
The device
is
intended for use in the electromagnetic environment specified below. The customer of the user of the device
should
assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The
device
use RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emission
CISPR 11
Class B
The
device
is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies