Holding down the MODE button for one-half second
cycles to the next system mode. These are comprised of
Meter-Counter Mode, Meter-Generator Mode,
Frequency Counter Mode, Power Meter Only Mode,
Signal Generator Mode, and RF Divider Mode.
1. Meter-Counter Mode: Displays both input power
level and frequency counter, with quick refresh rate and
low averaging. Up-down buttons have no function in
this mode.
2. Meter-Generator Mode: Displays the measured
input power level in dBm, and the user signal generator
frequency output in megahertz. Buttons adjust
frequency output.
3. Counter Only Mode: Displays the counted frequency
input in MHz with a longer gate time and more
precision in a large font. Mode button quick press saves
this reading as the power meter CAL frequency.
4. Meter Only Mode: Displays a more accurate power
level in dBm with more averaging and longer refresh
rate. Up-Down buttons set the CAL frequency.
5. Signal Generator Mode: Displays the frequency
output on the RF SMA port. Up/Down step buttons
change the frequency.
6. RF Divider Mode: Displays the current divider ratio
between the input frequency and the generator SMA
output port. Up/Down Step buttons change the divider
ratio (2, 4, or 8).
STANDALONE USAGE - SYSTEM MODES
TT7000
TT7000
– RF Meter / Counter / Generator - User Guide
www.dsinstruments.com