QBC STAR System Operator’s Manual
6000-300-000 5–2
6000-300-000 5-2
Parameter
Operating Ranges
Range Limits for
“Reference Results”
with Numerical
Results
Range Limits
for “Out of
Operating
Range” without
Numerical
Results
Granulocytes
0.8 – 70.0 x 109/L
0.6 -- 0.8 x 109/L
< 0.6 x 109/L
Lymph/Mono
0.8 – 99.9 x 109/L
0.3 -- 0.8 x 109/L
< 0.3 x 109/L
Platelets
20 -- 999 x 109/L
10 -- 20 x 109/L
< 10 x 109/L
WBC
1.6 – 99.9 x 109/L
0.9 – 1.6 x 109/L
< 0.9 x 109/L
% Granulocytes
Or
% Lymph/Mono
1% - 99%
If either Gr or LM is in this
range and if both Gr and LM
have numeric values
If either Gr or LM or
both is in this range
The table below listed the ranges between “Operating Ranges”, the ranges that numerical results will
be given and the ranges that numerical results will not be given.
Precision
Data on typical within-run precision tests on QBC STAR tubes tested in the QBC STAR system
are shown in the two tables below. The precision data represents the analysis of eleven whole
blood specimens, each assayed in replicates of 10.
Whole Blood Total Imprecision
Parameter
Mean Value
Mean %CV
HCT (%)
41.7
2.0 %
HB (g/dL)
14.0
1.9 %
PLT (x 10
9
/L)
235
6.0 %
WBC (x 10
9
/L)
6.0
6.4 %
Содержание QBC STAR
Страница 14: ...THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 28: ...THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 40: ...QBC STAR System Operator s Manual 6000 300 000 4 12 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 50: ...THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 72: ...6000 300 000 A 1 8 Appendices 6000 300 000 8 1...
Страница 74: ...THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 76: ...THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 78: ...THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 80: ...THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 86: ...THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 89: ...THIS PAGE INTENTIONALLY LEFT BLANK...