Control Panel Functions

Time Adjustment and Timer Operation:

 

The run time may be set from 30 seconds to 99 minutes 

and 30 seconds. Press the up and down arrow buttons next to the time display to adjust the run time. 
Adjustments may be made prior to a run or after a run begins. A quick tap will adjust the time by 30 
seconds. Hold down the button for 1 minute adjustments - continue to hold for 5 minute adjustments.

Press the down button once while 00:30 is displayed to access the “hold” feature. When a “hold” run 
is initiated, the timer will count up to keep track of total run time. The unit will run for a maximum of 99 
minutes and 30 seconds and will then shut off automatically. The “hold” run may also be terminated 
by pressing the STOP/OPEN button. The total run time during a “hold” run is displayed on the time 
display until the operator unlocks the lid by pressing the OPEN button or opens the lid (during the 
automatic unlock after the run ends).

Speed Adjustment / RCF Display:

 

Press the up and down arrow buttons next to the speed 

display to change to run speed. Adjustments may be made prior to a run or after a run begins. A quick 
tap will adjust the speed by 50 RPMs. Hold down the button for 100 RPM adjustments. Continue to 
hold for 500 RPM adjustments.

Press the RCF button at any time to convert the current speed setting to the corresponding force 
value (xg). The speed may be set by RCF by holding down the RCF button while using the up/down 
speed set buttons. The jumps made in the RCF value while holding down the RCF button and pressing 
the up and down arrow buttons correspond to 50 RPM adjustments in speed. Since the RCF is a 
function of the square of the speed, this jump size will vary.

Starting and Stopping a Run:

 With the lid switch closed, press the START button to begin 

a run. Press the STOP/OPEN button at any time during a run to terminate it.

Unlocking the Lid:

 

The lid is unlocked whenever the red UNLOCKED indicator light is illuminated. 

The lid unlocks automatically for 60 seconds at the end of a run. The lid can also be unlocked for an 
additional 15 seconds by pressing the OPEN/STOP button while the unit is idle. The lid cannot be 
unlocked while the rotor is spinning.

Imbalance Detection Reset:

 

This centrifuge is equipped with imbalance detection. If an imbalance 

is detected, the centrifuge will terminate the current run and will begin to brake to a stop. The words 
“BALANC” and “ERROR” will flash on the speed display. Once the rotor has stopped, open the lid 
the cancel the error reporting, balance the load and begin a new run. Alternately, the error reporting 
can be canceled by pressing the OPEN/STOP button

Advanced User Settings

This centrifuge allow for total control of the run by the user. You may adjust the acceleration rate, 
adjust the braking rate, adjust the imbalance sensitivity level, select from several audible settings, 
and use the optional countdown delay.

Acceleration Rate:

 

The user may adjust the acceleration rate from minimum (1) to maximum (10) 

using the following procedure:
1.  While the centrifuge is idle, press the PROGRAM button.
2.  Use the UP/DOWN arrow buttons next to the speed display until “ACCEL” is displayed in the 

speed display.

3.  The current acceleration rate is displayed in the time display. Use the ‘UP/ DOWN arrow buttons 

next to the time display to adjust the acceleration rate from 1 (minimum) to 10 (maximum).

4.  Press the PROGRAM button to exit or use the UP/DOWN arrow buttons next to the speed display 

to adjust additional settings.

Содержание 853VES

Страница 1: ...Laboratory Centrifuge Model 853VES Operator s Manual P N 7711012 Rev G MODEL 853VES IVD...

Страница 2: ...me Adjustment and Timer Operation 9 Speed Adjustment RCF Display 9 Starting and Stopping a Run 9 Unlocking the Lid 9 Advanced User Settings Acceleration Rate 9 Braking Rate 10 Countdown Delay 10 Imbal...

Страница 3: ...uch of a button High speed and force capability produces platelet poor free plasma and greatly reduces processing times The exclusive Drucker horizontal rotor allows for quick and easy sample loading...

Страница 4: ...Frequency Hz 50 60 Specifications General Specifications for the Horizon Model 853VES Centrifuge Contact your authorized dealer or Drucker Diagnostics to order accessories 75 mm Tube Holder p n 771304...

Страница 5: ...ge Buttons Running Indicator Light Latched Indicator Light Unlocked Indicator Light RPM Display Open Stop Button Start Button Memory Button Memory Preset Display Program Button Brake Display Up Down T...

Страница 6: ...tor light should be illuminated If it is not make sure that the lid is latched properly The centrifuge will not run unless the lid is latched properly and the LATCH indicator light is illuminated 7 Us...

Страница 7: ...D DURING A SPIN THAT REQUIRES THE CENTRIFUGE TO SHUT DOWN PRESS THE OPEN STOP BUTTON IMMEDIATELY 9 After time has elapsed the RUNNING indicator light will extinguish and the rotor will decelerate to a...

Страница 8: ...esigned to hold up to twelve 12 tube holders It comes with two tube holder types The purple holders are designed to hold tubes up to 10 mL up to 17 mm x 100 mm The blue holders are designed to hold tu...

Страница 9: ...ction of the square of the speed this jump size will vary Starting and Stopping a Run With the lid switch closed press the START button to begin a run Press the STOP OPEN button at any time during a r...

Страница 10: ...uge is equipped with imbalance detection This ensures that the centrifuge will safely shut down in the event that an imbalanced load is run or a malfunction occurs This feature is intended to provide...

Страница 11: ...layed 3 When the desired preset is found pause and the memory display will momentarily flash The stored configuration is now loaded Notes Once a configuration has been loaded it will remain active unt...

Страница 12: ...s or hand Remaining fragments may provide stress points on subsequent tubes and result in additional breakage If a tube breakage occurs carefully remove the tube holder Properly dispose of the sample...

Страница 13: ...ot run or an error message is displayed ERROR SPEED The centrifuge cannot reach full speed due to a problem with the rotor an inadequate power supply or other electrical problems Press the OPEN STOP b...

Страница 14: ...lid safety switch If this happens the centrifuge motor may lose power but the lid will still remain locked If the knob is accidentally moved and this situation should occur rotate the knob fully cloc...

Страница 15: ...ube Holder Cap p n 7713011 1 Cushion Spacer p n 1525 200 Shady Lane Suite 170 Philipsburg PA 16866 1 866 265 1486 U S only or 1 814 692 7661 Fax 814 692 7662 www druckerdiagnostics com INSTRUCTIONS FO...

Отзывы: