If any problems are found during the initial setup procedure, refer to the troubleshooting section. 
For further assistance, contact Drucker Diagnostics at +1-814-342-6205 or +1-814-692-7661.

1.   Plug the female end of the line cord into the rear of the centrifuge. Plug the male end into an 

approved electrical outlet. For electrical safety, the unit must always be properly grounded.

2.   For operator safety, the locking system is always active; requiring power and direction from the  

user to disengage it (the lid also automatically unlocks at the end of a run when it is safe to do 
so). To unlock the lid (in order to access the rotor chamber) press the ‘OPEN / STOP’ button 
on the control panel. The ‘UNLOCKED’ indicator light should illuminate. If it does not, refer to 
the section on troubleshooting. The lid will remain unlocked for 15 seconds after pressing the 
‘OPEN / STOP’ button.

3.  Turn the latch counterclockwise and open the lid.  
4.  Spin the rotor by hand; check for free and level rotation. If the rotor does not spin freely, refer 

to the section on troubleshooting.

5.  Place the test tube holders (or carriers depending on your rotor type) inside the rotor and verify 

that they are seated properly.

6.  Close the lid. Rotate the lid knob clockwise to its complete stop position. The ‘LATCHED’ indicator 

light should be illuminated. If it is not, make sure that the lid is latched properly. The centrifuge 
will not run unless the lid is latched properly and the ‘LATCH’ indicator light is illuminated.

7.  Use the up arrow button next to the RPM display to set the speed to full.
8.  Initiate a test run by pressing the ‘START’ button.  

(Continued on next page)

1. Unpack the centrifuge and verify that all of the supplied equipment is present.
2. Choose a setup location which meets the following criteria:

a) A bench top clearance height of 20” is required in order to open the lid.
b) The clearance envelope is the space around the centrifuge which is required for safety. Choose a 

setup location which will allow for a clearance envelope of at least 30” x 30”, (with the centrifuge 
at the center). No person or hazardous material shall be permitted in the clearance envelope 
during operation. The operator time within the envelope shall be limited to the time necessary 
for loading, unloading and centrifuge operation only. 

c) Proper ventilation is necessary to prevent the overheating of samples as well as premature 

failure of the centrifuge. Choose an area which will allow unencumbered air flow.

d) The centrifuge is designed to rest on its four rubber feet. No adjustment is necessary for leveling 

the centrifuge, however, the surface should be flat and level.

e) Be sure the outlet is always within reach as the line cord is the means of emergency disconnection.

Initial Setup Procedure

Setup Location

Содержание 853VES

Страница 1: ...Laboratory Centrifuge Model 853VES Operator s Manual P N 7711012 Rev G MODEL 853VES IVD...

Страница 2: ...me Adjustment and Timer Operation 9 Speed Adjustment RCF Display 9 Starting and Stopping a Run 9 Unlocking the Lid 9 Advanced User Settings Acceleration Rate 9 Braking Rate 10 Countdown Delay 10 Imbal...

Страница 3: ...uch of a button High speed and force capability produces platelet poor free plasma and greatly reduces processing times The exclusive Drucker horizontal rotor allows for quick and easy sample loading...

Страница 4: ...Frequency Hz 50 60 Specifications General Specifications for the Horizon Model 853VES Centrifuge Contact your authorized dealer or Drucker Diagnostics to order accessories 75 mm Tube Holder p n 771304...

Страница 5: ...ge Buttons Running Indicator Light Latched Indicator Light Unlocked Indicator Light RPM Display Open Stop Button Start Button Memory Button Memory Preset Display Program Button Brake Display Up Down T...

Страница 6: ...tor light should be illuminated If it is not make sure that the lid is latched properly The centrifuge will not run unless the lid is latched properly and the LATCH indicator light is illuminated 7 Us...

Страница 7: ...D DURING A SPIN THAT REQUIRES THE CENTRIFUGE TO SHUT DOWN PRESS THE OPEN STOP BUTTON IMMEDIATELY 9 After time has elapsed the RUNNING indicator light will extinguish and the rotor will decelerate to a...

Страница 8: ...esigned to hold up to twelve 12 tube holders It comes with two tube holder types The purple holders are designed to hold tubes up to 10 mL up to 17 mm x 100 mm The blue holders are designed to hold tu...

Страница 9: ...ction of the square of the speed this jump size will vary Starting and Stopping a Run With the lid switch closed press the START button to begin a run Press the STOP OPEN button at any time during a r...

Страница 10: ...uge is equipped with imbalance detection This ensures that the centrifuge will safely shut down in the event that an imbalanced load is run or a malfunction occurs This feature is intended to provide...

Страница 11: ...layed 3 When the desired preset is found pause and the memory display will momentarily flash The stored configuration is now loaded Notes Once a configuration has been loaded it will remain active unt...

Страница 12: ...s or hand Remaining fragments may provide stress points on subsequent tubes and result in additional breakage If a tube breakage occurs carefully remove the tube holder Properly dispose of the sample...

Страница 13: ...ot run or an error message is displayed ERROR SPEED The centrifuge cannot reach full speed due to a problem with the rotor an inadequate power supply or other electrical problems Press the OPEN STOP b...

Страница 14: ...lid safety switch If this happens the centrifuge motor may lose power but the lid will still remain locked If the knob is accidentally moved and this situation should occur rotate the knob fully cloc...

Страница 15: ...ube Holder Cap p n 7713011 1 Cushion Spacer p n 1525 200 Shady Lane Suite 170 Philipsburg PA 16866 1 866 265 1486 U S only or 1 814 692 7661 Fax 814 692 7662 www druckerdiagnostics com INSTRUCTIONS FO...