Advanced User Settings (continued)

Braking Rate:

 

The user may adjust the braking rate from minimum (slowest deceleration, 0) to 

maximum (quickest deceleration, 9) using the following procedure:
1. While the centrifuge is idle, press the PROGRAM button.
2. Use the UP/DOWN arrow buttons next to the speed display until “BRAKE” is displayed in the 

speed display.

3.  The current braking rate is displayed in the brake display. Use the UP/DOWN arrow buttons next 

to the brake display to adjust the braking rate from 0 (minimum) to 9 (maximum).

4.  Press the PROGRAM button to exit or use the UP/DOWN arrow buttons next to the speed display 

to adjust additional settings.

Countdown Delay:

 

When the countdown delay is turned on, the timer will not begin decrementing 

until the rotor has reached full speed as set by the user. This feature ensures that
the samples see full force for the total run time.
1. While the centrifuge is idle, press the PROGRAM button.
2. Use the UP/DOWN arrow buttons next to the speed display until “CNTDLY” is displayed in the 

speed display.

3. The current status of this feature (ON or OFF) is displayed in the time display. Press UP/DOWN 

arrows next to the time display to toggle the status between ON and OFF.

4.  Press the PROGRAM button to exit or use the UP/DOWN arrow buttons next to the speed display 

to adjust additional settings.

Imbalance Detection:

 

This centrifuge is equipped with imbalance detection. This ensures

that the centrifuge will safely shut down in the event that an imbalanced load is run or a malfunction
occurs. This feature is intended to provide for operator safety and to extend the life of the
centrifuge. The LOW setting will allow for a greater imbalance before shutting the centrifuge down.
1. While the centrifuge is idle, press the PROGRAM button.
2. Use the UP/DOWN arrow buttons next to the speed display until “BALANC” is
  displayed in the speed display.
3. The current status of this feature (LOW or HI) is displayed in the time display.
  Press UP/DOWN arrows next to the time display to toggle the status between
  LOW and HI.
4. Press the PROGRAM button to exit or use the UP/DOWN arrow buttons next to
  the speed display to adjust additional settings.

Audible Settings:

 

The user may select from six different audible indicator levels:

Setting  Button Press Beep?  End-of-run audible notification

1 .............NO  .............................. NONE
2 .............YES  ............................ NONE
3 .............NO  .............................. YES, 5 beeps
4 .............YES  ............................ YES, 5 beeps
5 .............NO  .............................. YES, continuous, user must press STOP/OPEN button to cancel
6 .............YES  ............................ YES, continuous, user must press STOP/OPEN button to cancel

 

1.  While the centrifuge is idle, press the PROGRAM button.

 

2.  Use the UP/DOWN arrow buttons next to the speed display until “BEEPER” is  

 

 

     displayed in the speed display.

 

3.  The current audible selection is displayed in the time display. Use the UP/ 

 

 

                   DOWN arrow buttons next to the time display to select a setting (1 to 6).
 

4.  Press the PROGRAM button to exit or use the UP/DOWN arrow buttons next to  

 

 

     the speed display to adjust additional settings.

Содержание 853VES

Страница 1: ...Laboratory Centrifuge Model 853VES Operator s Manual P N 7711012 Rev G MODEL 853VES IVD...

Страница 2: ...me Adjustment and Timer Operation 9 Speed Adjustment RCF Display 9 Starting and Stopping a Run 9 Unlocking the Lid 9 Advanced User Settings Acceleration Rate 9 Braking Rate 10 Countdown Delay 10 Imbal...

Страница 3: ...uch of a button High speed and force capability produces platelet poor free plasma and greatly reduces processing times The exclusive Drucker horizontal rotor allows for quick and easy sample loading...

Страница 4: ...Frequency Hz 50 60 Specifications General Specifications for the Horizon Model 853VES Centrifuge Contact your authorized dealer or Drucker Diagnostics to order accessories 75 mm Tube Holder p n 771304...

Страница 5: ...ge Buttons Running Indicator Light Latched Indicator Light Unlocked Indicator Light RPM Display Open Stop Button Start Button Memory Button Memory Preset Display Program Button Brake Display Up Down T...

Страница 6: ...tor light should be illuminated If it is not make sure that the lid is latched properly The centrifuge will not run unless the lid is latched properly and the LATCH indicator light is illuminated 7 Us...

Страница 7: ...D DURING A SPIN THAT REQUIRES THE CENTRIFUGE TO SHUT DOWN PRESS THE OPEN STOP BUTTON IMMEDIATELY 9 After time has elapsed the RUNNING indicator light will extinguish and the rotor will decelerate to a...

Страница 8: ...esigned to hold up to twelve 12 tube holders It comes with two tube holder types The purple holders are designed to hold tubes up to 10 mL up to 17 mm x 100 mm The blue holders are designed to hold tu...

Страница 9: ...ction of the square of the speed this jump size will vary Starting and Stopping a Run With the lid switch closed press the START button to begin a run Press the STOP OPEN button at any time during a r...

Страница 10: ...uge is equipped with imbalance detection This ensures that the centrifuge will safely shut down in the event that an imbalanced load is run or a malfunction occurs This feature is intended to provide...

Страница 11: ...layed 3 When the desired preset is found pause and the memory display will momentarily flash The stored configuration is now loaded Notes Once a configuration has been loaded it will remain active unt...

Страница 12: ...s or hand Remaining fragments may provide stress points on subsequent tubes and result in additional breakage If a tube breakage occurs carefully remove the tube holder Properly dispose of the sample...

Страница 13: ...ot run or an error message is displayed ERROR SPEED The centrifuge cannot reach full speed due to a problem with the rotor an inadequate power supply or other electrical problems Press the OPEN STOP b...

Страница 14: ...lid safety switch If this happens the centrifuge motor may lose power but the lid will still remain locked If the knob is accidentally moved and this situation should occur rotate the knob fully cloc...

Страница 15: ...ube Holder Cap p n 7713011 1 Cushion Spacer p n 1525 200 Shady Lane Suite 170 Philipsburg PA 16866 1 866 265 1486 U S only or 1 814 692 7661 Fax 814 692 7662 www druckerdiagnostics com INSTRUCTIONS FO...

Отзывы: