Laboratory Centrifuge
Model 853VES Operator’s Manual
P/N 7711012
Rev. G
MODEL 853VES
IVD
Страница 1: ...Laboratory Centrifuge Model 853VES Operator s Manual P N 7711012 Rev G MODEL 853VES IVD...
Страница 2: ...me Adjustment and Timer Operation 9 Speed Adjustment RCF Display 9 Starting and Stopping a Run 9 Unlocking the Lid 9 Advanced User Settings Acceleration Rate 9 Braking Rate 10 Countdown Delay 10 Imbal...
Страница 3: ...uch of a button High speed and force capability produces platelet poor free plasma and greatly reduces processing times The exclusive Drucker horizontal rotor allows for quick and easy sample loading...
Страница 4: ...Frequency Hz 50 60 Specifications General Specifications for the Horizon Model 853VES Centrifuge Contact your authorized dealer or Drucker Diagnostics to order accessories 75 mm Tube Holder p n 771304...
Страница 5: ...ge Buttons Running Indicator Light Latched Indicator Light Unlocked Indicator Light RPM Display Open Stop Button Start Button Memory Button Memory Preset Display Program Button Brake Display Up Down T...
Страница 6: ...tor light should be illuminated If it is not make sure that the lid is latched properly The centrifuge will not run unless the lid is latched properly and the LATCH indicator light is illuminated 7 Us...
Страница 7: ...D DURING A SPIN THAT REQUIRES THE CENTRIFUGE TO SHUT DOWN PRESS THE OPEN STOP BUTTON IMMEDIATELY 9 After time has elapsed the RUNNING indicator light will extinguish and the rotor will decelerate to a...
Страница 8: ...esigned to hold up to twelve 12 tube holders It comes with two tube holder types The purple holders are designed to hold tubes up to 10 mL up to 17 mm x 100 mm The blue holders are designed to hold tu...
Страница 9: ...ction of the square of the speed this jump size will vary Starting and Stopping a Run With the lid switch closed press the START button to begin a run Press the STOP OPEN button at any time during a r...
Страница 10: ...uge is equipped with imbalance detection This ensures that the centrifuge will safely shut down in the event that an imbalanced load is run or a malfunction occurs This feature is intended to provide...
Страница 11: ...layed 3 When the desired preset is found pause and the memory display will momentarily flash The stored configuration is now loaded Notes Once a configuration has been loaded it will remain active unt...
Страница 12: ...s or hand Remaining fragments may provide stress points on subsequent tubes and result in additional breakage If a tube breakage occurs carefully remove the tube holder Properly dispose of the sample...
Страница 13: ...ot run or an error message is displayed ERROR SPEED The centrifuge cannot reach full speed due to a problem with the rotor an inadequate power supply or other electrical problems Press the OPEN STOP b...
Страница 14: ...lid safety switch If this happens the centrifuge motor may lose power but the lid will still remain locked If the knob is accidentally moved and this situation should occur rotate the knob fully cloc...
Страница 15: ...ube Holder Cap p n 7713011 1 Cushion Spacer p n 1525 200 Shady Lane Suite 170 Philipsburg PA 16866 1 866 265 1486 U S only or 1 814 692 7661 Fax 814 692 7662 www druckerdiagnostics com INSTRUCTIONS FO...