Cycle start:
Press simultaneously and hold for a few seconds the -key and the
↑
-key.
↓
rStdL
The operator will be prompted to select if he wants to con nue the actual
condL
(
) or to start a new one (
/ previous stored data is lost).
cycle
condL
rStdL
On the display will appear cont1 or cont0. Select with the -key and
↑
the -key and press SET key. The acceptance of START whill be shown
↓
on the display by the presence of the icon . Each me a measurement point
is saved, the icon will flash for a second.
Cycle stop:
The cyle will stop automa cally at the set me t2.
Alterna vely hold down simultaneously, for a few seconds, the -key and
↑
the -key. The icon will be turned off.
↓
View data:
Press and hold simultaneously, for a few seconds, the SET-key and the -key.
↓
The acceptance of this se ng will be confirmed by flashing display of the icons
and . You can view all measurement points by using the ↑-key. To go
back one point, use the ↓-key. To exit this func on, press the SET-key.
In manual mode, you control the recording of the data by pressing the
PEAK-key.
7. Maintenance and cleaning
7.1 Maintenance
Maintenance shall be carried out by authorized personnel only. In daily use, the instrument
doesn‘t require maintenance. Repairs must be carried out only by the manufacturer.
CAUTION!
To avoid personal injury or damage to the calibrator, only use accessories supplied
by DRUCK & TEMPERATUR Leitenberger GmbH and ensure that no water finds it‘s
way into the housing.
7.2 Cleaning
CAUTION!
• Before cleaning, correctly disconnect the calibrator from the pressure supply,
switch it off and disconnect it from the mains.
• Clean the instrument with a moist cloth.
• Electrical connec ons must not come into contact with moisture.
• Clean the calibrator before returning them, in order to protect persons and the
environment from exposure to residual media.
• Residual media in a dismounted instrument can result in a risk to persons, the
environment and equipment. Take sufficient precau onary measures.