Using
indexes, you can find every single datalog capture point. Each index has a
XXXX
PPPP
pressure measurement value,
temperature value, and, in the case of manual mode,
me,
TTTT
which indicates the elapsed me since the sub-cycle started. With the addi on of par al sub-cycle
indexes and capture ming, in the case of automa c mode, you can resonstruct the ming of each
sub-cycle of datalog.
NOTE
Warning:
You can download only
You cannot download data while a datalog cycle is in progress.
the en re datalog. It is not possible to download the sub-cycle separately. Use the sub cycle index
to iden fy the start and up of each sub cycle.
The
,
,
commands return the current datalog se ngs and may differ
LB000000 L800000 L900000
from the informa on of the last datalog made, sent as a response to the
command.
L3000000
In the response to command
, the string
can range from a minimum
L600000
XXXXPPPPTTTT
of 8 bytes to a maximum of 16 bytes,
is present only if the temperature was recorded in the
TTTT
capture cycle,
is present only if the capture cycle was manual.