For informaon on returning the calibrator and its sensor(s)
see chapter "9.2 Return".
7.3 Recalibraon
DKD DA
/
kkS er icate -
Cer icates:
C
f
s Factory
f
We recommend that the instrument is regularly recalibrated by the manufacturer, with
me intervals of approx. 12 months. Every factory recalibraon includes, addionally, an
extensive free-of-charge check of all system parameters with respect to their compliance
with the speciicaon. The basic se ngs will be corrected if necessary.
Display
Causes
Measures
low BAT
The rechargeable ba ery's capacity is
less than 10 %.
Recharge the internal rechargeable
ba ery using the ba ery charger.
Sensor
Overlow
The current pressure value is outside
the measuring range of the model
LPC-2
reference pressure sensor
Adjust the pressure immediately and
appropriately.
No display
or strange
symbols, device
is not respond-
ing to input
Malfuncon during operaon.
Switch of the instrument and switch
on again aer 5 seconds.
No display
The instrument is not detected by the
PC.
Check whether the
interface
USB
is connected correctly and
is
USB
acvated in the
(see chapter
calibrator
6.8.2)
No display
When switching the instrument on,
aer 10 seconds no measured value
is displayed, but instead, the whole
display is dark.
Recharge the internal rechargeable
ba ery using the ba ery charger.
The display is dark and the recharge-
able ba ery cannot be charged via the
ba ery charger.
Check that the ba ery charger is
connected properly, and also check
(using appropriately quali ied person
f
-
nel) that the power supply voltage is
correct.
8. Faults