4.4.1 Connecon of the
reference pressure sensor
LPC-2
CAUTION!
Only use model
reference pressure sensors! Using other
LPC-2
sensors could damage both the
200
200-T
LPC
LPC
/
calibrator
LR-
Cal
and the reference pressure sensors.
Switch the
o f before changing sensors.
calibrator f
Before switching the instrument on, connect the sensor, otherwise it
may not be correctly iden ied by the instrument. The iden icaon of
f
f
the
LPC-2
reference pressure sensors can take up to 60 seconds.
For gauge or relave pressure sensors, there is a pressure-equalising
vent in the top of the sensor under the plasc cover. This vent (with
integrated membrane) must remain clear, without fail!
CAUTION!
Only ever use the original
connecng cable in the
LEITENBERGER
operaon of
LPC-2
reference pressure sensors.
The
and the reference pressure sensor are connected to each other
calibrator
electrically using a separate connecng cable. When exchanging the sensor, use the
8-pin plug connector at the sensor.
Connecon between
instrument and sensor
via connecng cable
Reference pressure
sensor
LPC-2
Calibrator
or
200
LPC
200-T
LPC
LR-
Cal
LR-
Cal