background image

 - 13 -

4. INSPECTION AND FUNCTIONAL TESTING 

Safety measures as described in Section 1 must be observed when inspecting 
or testing a chair. Only authorized caregivers or maintenance staff should 
attempt to inspect or test a chair. 

4.1 Inspection 

W

e recommend regular visual inspection for signs of wear, damage, loose 

or missing fittings, and other safety concerns. If a breakage, defect, or 
operational problem is detected, the chair must be repaired inspected and 
tested for function before it is returned to service. 

W

e recommend that the chair be inspected as often as each use if there 

is any reason to be concerned about the possibility of increased wear or 
loose or missing fittings. 

At a minimum, in regular use the chair should be 

inspected on a bimonthly basis.

 The visual inspection procedure should 

include at least the observation of all the fittings (fasteners): 

T

he visual inspection procedure should include at least the observation of 

the following parts: 

1.

   The cylinders that position the seat tilt, back recline, and independent  

 

leg rest elevation. 

2.  The vinyl straps used in the seat, back, leg rest, footrest and wings 
3.  The cushions 
4.  The black handles and cables. 
5.  The armrest height adjustment pins 

T

he visual inspection procedure should include the observation of any 

installed accessories. The visual inspection should be performed by the 
facility, or if in a private residence, by the individual responsible for the chair. 

4.2 Functional Testing 

We recommend that the chair should be tested for operation of the chair’s 
functions without a resident in the chair. 

The testing may be as often as each 

use if there is any reason to be concerned about the possibility of increased 
wear or damage to the chair’s components. At a minimum, in regular use the 
chair should be tested for functions as described in Section 3. 

If

 the caregiver or maintenance department performing the functional 

testing believes that any function is not operating correctly, the chair should 
be taken out of service until a satisfactory functional test can be completed. 

The caregiver or maintenance department performing the testing should 
be aware that the seat tilt, back recline, and footrest (leg rest) elevation 
operations will be more difficult without a resident in the chair. 

W

hen performing repairs or maintenance, do not use lubricants that contain 

solvents. Solvents will damage many of the moving components in the chair. 
If necessary, a white, food grade grease (lubricant) may be used on the 
sliding components in the chair. Do not use spray lubricants on any part of 
the chair.

Содержание CLN400-20

Страница 1: ...Full Tilt and Recliner Item CLN400 20 Distributor Rep This manual must be given to the caregiver responsible for this chair and its occupant Caregiver Before using this chair read this manual thoroug...

Страница 2: ...sident s Skin 6 2 6 Improper Use 7 2 6 1 Improper Restraint Use Risk of Serious Injury 7 2 7 Cleaning 8 2 8 Maintenance 9 2 9 Resident Specific Instructions 9 3 OPERATION AND MOVEMENT 10 3 1 Seat Tilt...

Страница 3: ...050 tel 516 998 4600 website www drivemedical com FOREWORD This document provides guidance on the safe and effective operation Information in this manual must be followed at all times Anyone involved...

Страница 4: ...or the chair used by the resident and may include the resident s family members or guardian Seat Tilt refers to changing the relative angle between the chair s seat and the chair frame or ground witho...

Страница 5: ...o ensure the overall safety comfort and well being of the resident caregiver and third party All operations and adjustments required for the resident should be determined by the resident s primary car...

Страница 6: ...visual inspection and before use These obligations apply to the chair s first use and to all subsequent uses Section 4 Inspection and Functional Testing 2 5 Hazards 2 5 1 Position of Chair Danger of...

Страница 7: ...ace to minimize the risk of tipping over 2 5 4 Total Lock Wheel Brakes Danger of Falling The special casters found on the DRIVE chair have total lock brakes which prevent the wheels from turning and s...

Страница 8: ...be required to ensure the safety of these residents during these operations 2 5 6 Unintended Movement Danger of Falling or Collision We recommend DRIVE chairs for indoor use and where there is not eno...

Страница 9: ...mmend a thigh belt be used In all cases it is the resident s primary caregiver who must take responsibility for the safety of the resident if restraints are used You should NOT have 1 Unauthorized ope...

Страница 10: ...brands include Lysol and Clorox home and Institutional use PDI healthcare setting Sani Professional food service and janitorial setting Sanitization Hand hygiene is of critical importance throughout t...

Страница 11: ...or sloped surfaces 4 Any unauthorized person Do not use lubricants that contain solvents Solvents may compromise any rubber components on the chair If necessary white food grade grease or lubricant c...

Страница 12: ...rt and assist with daily care activities WARNING Always tilt slightly before Recliner to prevent resident from sliding To Activate the Back Recline Place your LEFT hand on the top bar to assist moveme...

Страница 13: ...sters which provide upper body lateral support to the resident To Completely Remove the Wing Depress the top button 1 and pull the wing from the receptacle To Insert the Wing Insert the lower end of t...

Страница 14: ...e panel 3 9 Adjustable Seat Height To adjust the seat height Using 7 16 wrenches remove the bolts Raise or lower the front of the seat to correspond with the holes at the desired seat height Insert th...

Страница 15: ...height adjustment pins The visual inspection procedure should include the observation of any installed accessories The visual inspection should be performed by the facility or if in a private residen...

Страница 16: ...25 8 655mm I Arm pad length 14 356mm J Arm pad width 2 87 73mm k Seat depth 17 7 450mm L Back height 32 6 830mm M Width between two shoulder rest 13 330mm N Min area size between two shoulder pillow...

Страница 17: ...2 G2 Back angle 53 Recline Item No Description Spec size inch mm A1 Overall height 23 4 595mm B1 Overall length 65 2 1656mm D1 Seat height 21 6 550mm E1 Foot pad height 22 0 560mm F1 Seat angle 5 G1 B...

Страница 18: ...ROUBLESHOOTING AND REPAIR Symptom Problem Solution The chair is not Recliner or tilting or leg rest is not moving when the handle is squeezed The pin in the end of the cylinder is not being depressed...

Страница 19: ...an expression of our confidence in the mate rials and workmanship of our products and our assur ance to the consumer of years of dependable service This warranty does not cover device failure due to...

Страница 20: ...n NY 11050 Phone 516 998 4600 Fax 516 998 4601 www drivemedical com VER A 4 21 UDEM International Certification Auditing Training Centre Industry and Trade Inc Co Mutlukent Mahallesi 2073 Sokak Eski 9...

Отзывы: